摘要
Busulfan is widely used in bone marrow transplantation. Increased area under the plasma concentration curve has been shown to correlate with venoocclusive disease, which occurs in similar to 20% of these patients. The authors developed a high performance liquid chromatography assay for the determination of plasma busulfan concentration using ultraviolet detection and a single-step derivatization and extraction. The absolute retention times of busulfan and the internal standard were 2.8 and 5.6 min, respectively. The assay possessed linearity up to 200 mu mol/l, sensitivity to at least 0.2 mu mol/l, average recovery of 101%, and run-to-run precision (n = 34) of <7%. Furthermore, the assay proved to be free of interference from 59 medications and correlated highly with a validated method employing gas chromatography with electron capture detection (slope = 0.90, intercept = 0.17, r = 0.98, n = 31). The authors conclude that the method described here is ideally suited for the therapeutic monitoring of busulfan.
引用
收藏
页码:169 / 174
页数:6
相关论文
共 17 条
[1]
PRETRANSPLANT CONDITIONING WITH BUSULFAN (MYLERAN) AND CYCLOPHOSPHAMIDE FOR NONMALIGNANT DISEASES - ASSESSMENT OF ENGRAFTMENT FOLLOWING HISTOCOMPATIBLE ALLOGENEIC BONE-MARROW TRANSPLANTATION [J].
BLAZAR, BR ;
RAMSAY, NKC ;
KERSEY, JH ;
KRIVIT, W ;
ARTHUR, DC ;
FILIPOVICH, AH .
TRANSPLANTATION, 1985, 39 (06) :597-603
[2]
DETERMINATION OF BUSULFAN IN HUMAN-PLASMA BY GAS-CHROMATOGRAPHY WITH ELECTRON-CAPTURE DETECTION [J].
CHEN, TL ;
GROCHOW, LB ;
HUROWITZ, LA ;
BRUNDRETT, RB .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1988, 425 (02) :303-309
[3]
EHRSSON H, 1983, J PHARM SCI, V72, P1203, DOI 10.1002/jps.2600721024
[4]
HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF BUSULFAN IN HUMAN SERUM WITH ONLINE DERIVATIZATION, COLUMN-SWITCHING AND ULTRAVIOLET ABSORBENCY DETECTION [J].
FUNAKOSHI, K ;
YAMASHITA, K ;
CHAO, WF ;
YAMAGUCHI, M ;
YASHIKI, T .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1994, 660 (01) :200-204
[5]
PHARMACOKINETICS OF BUSULFAN - CORRELATION WITH VENO-OCCLUSIVE DISEASE IN PATIENTS UNDERGOING BONE-MARROW TRANSPLANTATION [J].
GROCHOW, LB ;
JONES, RJ ;
BRUNDRETT, RB ;
BRAINE, HG ;
CHEN, TL ;
SARAL, R ;
SANTOS, GW ;
COLVIN, OM .
CANCER CHEMOTHERAPY AND PHARMACOLOGY, 1989, 25 (01) :55-61
[6]
GROCHOW LB, 1993, SEMIN ONCOL, V20, P18
[7]
GAS-CHROMATOGRAPHIC DETERMINATION OF BUSULFAN IN PLASMA WITH ELECTRON-CAPTURE DETECTION [J].
HASSAN, M ;
EHRSSON, H .
JOURNAL OF CHROMATOGRAPHY, 1983, 277 (OCT) :374-380
[8]
MEASUREMENT OF BUSULFAN IN PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
HENNER, WD ;
FURLONG, EA ;
FLAHERTY, MD ;
SHEA, TC ;
PETERS, WP .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1987, 416 (02) :426-432
[9]
HOBBS JR, 1986, BONE MARROW TRANSPL, V1, P201
[10]
VENOOCCLUSIVE DISEASE OF THE LIVER FOLLOWING BONE-MARROW TRANSPLANTATION [J].
JONES, RJ ;
LEE, KSK ;
BESCHORNER, WE ;
VOGEL, VG ;
GROCHOW, LB ;
BRAINE, HG ;
VOGELSANG, GB ;
SENSENBRENNER, LL ;
SANTOS, GW ;
SARAL, R .
TRANSPLANTATION, 1987, 44 (06) :778-783