Prevention of tracheal aspiration using the pressure-limited tracheal tube cuff

被引:33
作者
Young, PJ [1 ]
Basson, C [1 ]
Hamilton, D [1 ]
Ridley, SA [1 ]
机构
[1] Norfolk & Norwich Hosp, Intens Care Unit, Norwich NR1 3SR, Norfolk, England
关键词
equipment; tracheal tube; complications; aspiration; pneumonia; ventilation; mechanical; intensive care;
D O I
10.1046/j.1365-2044.1999.00850.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A new design of tracheal tube cuff, the pressure-limited cuff, used with a constant-pressure inflation system, was compared with a high-volume low-pressure cuffed tracheal tube for leakage of dye placed in the subglottic space into the trachea. Patients requiring ventilation on the intensive cars unit were randomly allocated into two groups, one for each type of cuff, and blue food dye was instilled daily via a fine catheter above the cuff into the subglottic space. There were eight patients in the high-volume low-pressure group and seven in the pressure-limited cuff group. Dye leaked into the trachea in seven (87%) of the high-volume low-pressure group compared with none (0%) of the pressure-limited cuff group (p < 0.01). This study demonstrates that the pressure-limited cuffed tracheal tube, in combination with a constant-pressure inflation device, prevents leakage of fluid into the lungs that occurs with high-volume low-pressure cuffs in the critically ill, intubated patient.
引用
收藏
页码:559 / 563
页数:5
相关论文
共 14 条
[1]  
BONTEN MJM, 1994, CRIT CARE MED, V22, P1683
[2]   PREVENTION OF HOSPITAL-ACQUIRED PNEUMONIA - MEASURING EFFECT IN OUNCES, POUNDS, AND TONS [J].
CRAVEN, DE .
ANNALS OF INTERNAL MEDICINE, 1995, 122 (03) :229-231
[3]   THE PATHOGENESIS OF VENTILATOR-ASSOCIATED PNEUMONIA .1. MECHANISMS OF BACTERIAL TRANSCOLONIZATION AND AIRWAY INOCULATION [J].
ESTES, RJ ;
MEDURI, GU .
INTENSIVE CARE MEDICINE, 1995, 21 (04) :365-383
[4]   NOSOCOMIAL PNEUMONIA IN VENTILATED PATIENTS - A COHORT STUDY EVALUATING ATTRIBUTABLE MORTALITY AND HOSPITAL STAY [J].
FAGON, JY ;
CHASTRE, J ;
HANCE, AJ ;
MONTRAVERS, P ;
NOVARA, A ;
GIBERT, C .
AMERICAN JOURNAL OF MEDICINE, 1993, 94 (03) :281-288
[5]   DESIGN AND DEVELOPMENT OF ULTRATHIN-WALLED, NONKINKING ENDOTRACHEAL-TUBES OF A NEW NO-PRESSURE LARYNGEAL SEAL DESIGN - A PRELIMINARY-REPORT [J].
KOLOBOW, T ;
TSUNO, K ;
ROSSI, N ;
APRIGLIANO, M .
ANESTHESIOLOGY, 1994, 81 (04) :1061-1067
[6]   PREVENTION OF NOSOCOMIAL PNEUMONIA IN INTUBATED PATIENTS - RESPECTIVE ROLE OF MECHANICAL SUBGLOTTIC SECRETIONS DRAINAGE AND STRESS-ULCER PROPHYLAXIS [J].
MAHUL, P ;
AUBOYER, C ;
JOSPE, R ;
ROS, A ;
GUERIN, C ;
ELKHOURI, Z ;
GALLIEZ, M ;
DUMONT, A ;
GAUDIN, O .
INTENSIVE CARE MEDICINE, 1992, 18 (01) :20-25
[7]   FAILURE OF A HIGH-COMPLIANCE LOW-PRESSURE CUFF TO PREVENT ASPIRATION [J].
PAVLIN, EG ;
VANNIMWEGEN, D ;
HORNBEIN, TF .
ANESTHESIOLOGY, 1975, 42 (02) :216-219
[8]   COMPARISON OF BLUE-DYE VISUALIZATION AND GLUCOSE-OXIDASE TEST STRIP METHODS FOR DETECTING PULMONARY ASPIRATION OF ENTERAL FEEDINGS IN INTUBATED ADULTS [J].
POTTS, RG ;
ZAROUKIAN, MH ;
GUERRERO, PA ;
BAKER, CD .
CHEST, 1993, 103 (01) :117-121
[9]   New ultrathin-walled endotracheal tube with a novel laryngeal seal design: Long-term evaluation in sheep [J].
Robotham, JL .
ANESTHESIOLOGY, 1996, 84 (01) :A27-A27
[10]   ASPIRATION BEYOND ENDOTRACHEAL CUFFS [J].
SEEGOBIN, RD ;
VANHASSELT, GL .
CANADIAN ANAESTHETISTS SOCIETY JOURNAL, 1986, 33 (03) :273-279