Accuracy of three electronic monitors for metered-dose inhalers

被引:62
作者
Julius, SM
Sherman, JM
Hendeles, L
机构
[1] Univ Florida, Pediat Pulm Div, Gainesville, FL USA
[2] Univ Florida, Coll Pharm, Gainesville, FL USA
关键词
adherence; device; metered-dose inhaler;
D O I
10.1378/chest.121.3.871
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Prior studies indicate that some devices used to monitor metered-dose inhaler (MDI) use are not accurate. Objective: To assess the accuracy of the Doser CT (NEW-MED Corporation; Waltham, MA), the MDILog (Medtrac Technologies; Lakewood, CO), and the SmartMist (Aradigm Corporation; Hayward, CA) in a bench-top study. Design: One, two, and four puffs of fluticasone propionate NIDI (Flovent; GlaxoWellcome; Research Triangle Park, NC) were actuated twice daily for 30 days with two units of each device. The date and time of each actuation were recorded in a log and then compared with the output of the device. The percentage of doses recorded accurately was compared by, analysis of variance. Measurements and results: The SmartMist was 100% accurate, while both the Doser CT and MDILog devices occasionally recorded spurious actuations. The Doser CT also had missed actuations after the counter had prematurely, reached zero secondary, to the spurious recordings. The accuracy (mean +/-SD) was 94.3 +/- 2.9% for the Doser CT and 90.1 +/- 6.9% for the MDILog (p = 0.21). The dose regimen actuated and duration of use did not significantly, affect accuracy. Conclusion: All three devices are sufficiently accurate to monitor adherence in most clinical settings.
引用
收藏
页码:871 / 876
页数:6
相关论文
共 16 条
  • [1] INVESTIGATION OF A CLUSTER OF DEATHS OF ADOLESCENTS FROM ASTHMA - EVIDENCE IMPLICATING INADEQUATE TREATMENT AND POOR PATIENT ADHERENCE WITH MEDICATIONS
    BIRKHEAD, G
    ATTAWAY, NJ
    STRUNK, RC
    TOWNSEND, MC
    TEUTSCH, S
    [J]. JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1989, 84 (04) : 484 - 491
  • [2] Reliability of the Medtrac MDI Chronolog
    Brueckner, JW
    Marshik, P
    Sherman, J
    Hendeles, L
    [J]. JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1997, 100 (04) : 488 - 491
  • [3] MEASURING COMPLIANCE WITH INHALED MEDICATION IN ASTHMA
    COUTTS, JAP
    GIBSON, NA
    PATON, JY
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD, 1992, 67 (03) : 332 - 333
  • [4] METERED-DOSE INHALER USAGE IN SUBJECTS WITH ASTHMA - COMPARISON OF NEBULIZER CHRONOLOG AND DAILY DIARY RECORDINGS
    GONG, H
    SIMMONS, MS
    CLARK, VA
    TASHKIN, DP
    [J]. JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1988, 82 (01) : 5 - 10
  • [5] MAWHINNEY H, 1991, ANN ALLERGY, V66, P294
  • [6] Noncompliance and treatment failure in children with asthma
    Milgrom, H
    Bender, B
    Ackerson, L
    Bowry, P
    Smith, B
    Rand, C
    [J]. JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1996, 98 (06) : 1051 - 1057
  • [7] IMPROVING INHALER ADHERENCE IN A CLINICAL-TRIAL THROUGH THE USE OF THE NEBULIZER CHRONOLOG
    NIDES, MA
    TASHKIN, DP
    SIMMONS, MS
    WISE, RA
    LI, VC
    RAND, CS
    [J]. CHEST, 1993, 104 (02) : 501 - 507
  • [8] PATTERSON R, 1991, ANN ALLERGY, V67, P138
  • [9] METERED-DOSE INHALER ADHERENCE IN A CLINICAL-TRIAL
    RAND, CS
    WISE, RA
    NIDES, M
    SIMMONS, MS
    BLEECKER, ER
    KUSEK, JW
    LI, VC
    TASHKIN, DP
    [J]. AMERICAN REVIEW OF RESPIRATORY DISEASE, 1992, 146 (06): : 1559 - 1564
  • [10] A CASE-CONTROL STUDY OF DEATHS FROM ASTHMA
    REA, HH
    SCRAGG, R
    JACKSON, R
    BEAGLEHOLE, R
    FENWICK, J
    SUTHERLAND, DC
    [J]. THORAX, 1986, 41 (11) : 833 - 839