A hollow cathode proton transfer reaction time of flight mass spectrometer

被引:49
作者
Ennis, CJ [1 ]
Reynolds, JC [1 ]
Keely, BJ [1 ]
Carpenter, LJ [1 ]
机构
[1] Univ York, Dept Chem, York YO10 5DD, N Yorkshire, England
关键词
proton transfer reaction mass spectrometry; atmospheric analysis; time of flight mass spectrometry;
D O I
10.1016/j.ijms.2005.09.008
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
A proton transfer reaction mass spectrometer has been developed that couples a readily portable hollow cathode proton transfer reaction cell with a reflectron time of flight detector (PTR-ToF). The instrument is capable of detection of trace gas components as low as 1 ppbv on a timescale of between 10 and 60 s and with a sensitivity of 3.7 ncps ppbv(-1) (normalised counts per second per parts per billion by volume) for toluene and 28 ncps ppbv(-1) for acetone. The PTR-ToF instrument developed is capable of high mass resolution, high mass accuracy (similar to 300 ppm) and multiplexed spectral acquisition, leading to rapid and selective online analysis of trace components in complex gas mixtures. Attention to the materials chemistry of the system and the reaction conditions in the PTR cell are expected to enable the detection limit of this instrument to be lowered and the sensitivity enhanced. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:72 / 80
页数:9
相关论文
共 19 条
[1]   Demonstration of proton-transfer reaction time-of-flight mass spectrometry for real-time analysis of trace volatile organic compounds [J].
Blake, RS ;
Whyte, C ;
Hughes, CO ;
Ellis, AM ;
Monks, PS .
ANALYTICAL CHEMISTRY, 2004, 76 (13) :3841-3845
[2]   SIMION for the personal computer in reflection [J].
Dahl, DA .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2000, 200 (1-3) :3-25
[3]   Sensitivity and specificity of atmospheric trace gas detection by proton-transfer-reaction mass spectrometry [J].
de Gouw, J ;
Warneke, C ;
Karl, T ;
Eerdekens, G ;
van der Veen, C ;
Fall, R .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2003, 223 (1-3) :365-382
[4]  
Hansel A, 1998, RAPID COMMUN MASS SP, V12, P871, DOI 10.1002/(SICI)1097-0231(19980715)12:13<871::AID-RCM245>3.0.CO
[5]  
2-L
[6]  
HANSEL A, 2004, INT J MASS SPECTROM, V239, P77
[7]   Performance characteristics and applications of a proton transfer reaction-mass spectrometer for measuring volatile organic compounds in ambient air [J].
Hayward, S ;
Hewitt, CN ;
Sartin, JH ;
Owen, SM .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2002, 36 (07) :1554-1560
[8]   The application of proton transfer reaction-mass spectrometry (PTR-MS) to the monitoring and analysis of volatile organic compounds in the atmosphere [J].
Hewitt, CN ;
Hayward, S ;
Tani, A .
JOURNAL OF ENVIRONMENTAL MONITORING, 2003, 5 (01) :1-7
[9]  
HOWATSON AM, 1976, INTRO FAS DISCHARGE
[10]  
Ikezoe Y., 1987, Gas Phase Ion-Molecule Reaction Rate Constants through 1986