Rotational sensors-a comparison of different sensor types

被引:70
作者
Bernauer, Felix [1 ]
Wassermann, Joachim [1 ]
Igel, Heiner [1 ]
机构
[1] Univ Munich, Dept Earth & Environm Sci, D-80333 Munich, Germany
关键词
Rotational seismology; Instrumentation; Rotational motion sensors; Absolute calibration; Instrument self noise; SEISMOLOGY; SPECTRA; MOTION; OBSPY; NOISE;
D O I
10.1007/s10950-012-9286-7
中图分类号
P3 [地球物理学]; P59 [地球化学];
学科分类号
0708 ; 070902 ;
摘要
We present laboratory tests for two types of rotational motion sensors, the liquid-based rotational seismometers type R1 and type R2 manufactured by Eentec and the closed loop fiber optic gyroscope LCG-Demonstrator by Northrop Grumman LITEF. All instruments were calibrated absolutely at different temperatures, characterization and quantification of self-noise was carried out, and a comparison with the ring laser gyroscope G from the Geodetic Observatory in Wettzell, Germany is drawn. The generator constant of the R1 varies up to 27% in the nominal operating temperature range. In the closed-loop system LCG-Demonstrator, the compensation for temperature variation works very well, and the generator constant can be seen as constant within the error bars. For both instrument types, we measured sensitivities in the order of 10 (-aEuro parts per thousand 7) rad/s in a period range from 10 to 100 s. While this sensitivity is already sufficient for civil engineering applications, it has to be improved by at least 1 order of magnitude for applications in weak motion seismology.
引用
收藏
页码:595 / 602
页数:8
相关论文
共 16 条
[1]   STATISTICS OF ATOMIC FREQUENCY STANDARDS [J].
ALLAN, DW .
PROCEEDINGS OF THE INSTITUTE OF ELECTRICAL AND ELECTRONICS ENGINEERS, 1966, 54 (02) :221-&
[2]  
[Anonymous], 2006, ROTATIONAL MOTIONS S
[3]  
Barnes JA, 1970, 394 NAT BUR STAND
[4]   ObsPy: A Python']Python Toolbox for Seismology [J].
Beyreuther, Moritz ;
Barsch, Robert ;
Krischer, Lion ;
Megies, Tobias ;
Behr, Yannik ;
Wassermann, Joachim .
SEISMOLOGICAL RESEARCH LETTERS, 2010, 81 (03) :530-533
[5]  
Eentec, 2006, EL SENS TRANSD
[6]   Method for Calculating Self-Noise Spectra and Operating Ranges for Seismographic Inertial Sensors and Recorders [J].
Evans, J. R. ;
Followill, F. ;
Hutt, C. R. ;
Kromer, R. P. ;
Nigbor, R. L. ;
Ringler, A. T. ;
Steim, J. M. ;
Wielandt, E. .
SEISMOLOGICAL RESEARCH LETTERS, 2010, 81 (04) :640-646
[7]   Tilts in strong ground motion [J].
Graizer, Vladimir .
BULLETIN OF THE SEISMOLOGICAL SOCIETY OF AMERICA, 2006, 96 (06) :2090-2102
[8]   Strong Motion Recordings and Residual Displacements: What Are We Actually Recording in Strong Motion Seismology? [J].
Graizer, Vladimir .
SEISMOLOGICAL RESEARCH LETTERS, 2010, 81 (04) :635-639
[9]   Rotational motions induced by the M8.1 Tokachi-oki earthquake, September 25, 2003 [J].
Igel, H ;
Schreiber, U ;
Flaws, A ;
Schuberth, BSA ;
Velikoseltsev, A ;
Cochard, A .
GEOPHYSICAL RESEARCH LETTERS, 2005, 32 (08) :1-5
[10]   Recording Rotational and Translational Ground Motions of Two TAIGER Explosions in Northeastern Taiwan on 4 March 2008 [J].
Lin, Chin-Jen ;
Liu, Chun-Chi ;
Lee, William H. K. .
BULLETIN OF THE SEISMOLOGICAL SOCIETY OF AMERICA, 2009, 99 (2B) :1237-1250