Effect of sequence length, sequence frequency, and data acquisition rate on the performance of a Hadamard transform time-of-flight mass spectrometer

被引:28
作者
Fernández, FM
Vadillo, JM
Engelke, F
Kimmel, JR
Zare, RN [1 ]
Rodriguez, N
Wetterhall, M
Markides, K
机构
[1] Stanford Univ, Dept Chem, Stanford, CA 94305 USA
[2] Pacific NW Natl Lab, ESML, Richland, WA 99352 USA
[3] Uppsala Univ, Dept Chem, Uppsala, Sweden
关键词
D O I
10.1016/S1044-0305(01)00322-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Various factors influencing the performance of a Hadamard transform time-of-flight mass spectrometer (HT-TOFMS) have been investigated. Using a nitrogen corona discharge to produce an ion stream of N-2(+), N-3(+), and N-4(+), it is found for spectra containing only N-4(+) that the signal-to-noise ratio (SNR) closely approaches the value calculated from the ion background by assuming that the ion background follows a Poisson distribution. In contrast, for a more intense beam containing N-2(+), N-3(+), and N-4(+), the SNR is less than its theoretical value because of the appearance of discrete spikes in the mass spectrum caused by deviations in the actual modulation sequence from the ideal one. These spikes can be reduced, however, by decreasing the modulation voltage. Under these optimized conditions, the pseudo-random sequence length is varied to understand how it alters SNR, mass resolution, and scan speed. When the length of the pseudo-random sequence is doubled, the SNR increases by root2 while the time necessary to record a mass spectrum also doubles. Mass resolution can be varied between 500 and 1200 at m/z = 609 as the sequence length, modulation speed (10 MHz, 25 MHz), and acquisition rate (up to 50 MHz) are changed. Scan speeds of 6000 passes per s can be obtained using a sequence containing 4095 elements modulated at 25 MHz. The capability to tailor the HT-TOFMS to increase the scan speed and resolution with a constant 50% duty cycle makes the technique extremely appealing as a mass analyzer for measuring rapid changes in the composition of an ion stream. (C) 2001 American Society for Mass Spectrometry.
引用
收藏
页码:1302 / 1311
页数:10
相关论文
共 23 条
[1]  
Barnidge DR, 1999, RAPID COMMUN MASS SP, V13, P994, DOI 10.1002/(SICI)1097-0231(19990615)13:11<994::AID-RCM596>3.0.CO
[2]  
2-W
[3]   Hadamard transform time of flight mass spectrometry [J].
Brock, A ;
Rodriguez, N ;
Zare, RN .
ANALYTICAL CHEMISTRY, 1998, 70 (18) :3735-3741
[4]   Characterization of a Hadamard transform time-of-flight mass spectrometer [J].
Brock, A ;
Rodriguez, N ;
Zare, RN .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2000, 71 (03) :1306-1318
[5]   THE DESIGN AND PERFORMANCE OF AN ION-TRAP STORAGE REFLECTRON TIME-OF-FLIGHT MASS-SPECTROMETER [J].
CHIEN, BM ;
MICHAEL, SM ;
LUBMAN, DM .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1994, 131 :149-179
[6]  
Cotter R. J, 1994, ACS S SERIES, V549
[7]   TRANSLATIONAL ENERGY-DEPENDENCE OF AR(+)+H2-]ARX(+)+Y, AR(+)+D2-]ARX(+)+Y, AR(+)+HD-]ARX(+)+Y FROM THERMAL TO 30 EV CM [J].
ERVIN, KM ;
ARMENTROUT, PB .
JOURNAL OF CHEMICAL PHYSICS, 1985, 83 (01) :166-189
[8]  
Guilhaus M, 2000, MASS SPECTROM REV, V19, P65, DOI 10.1002/(SICI)1098-2787(2000)19:2<65::AID-MAS1>3.0.CO
[9]  
2-E
[10]   Masking, photobleaching, and spreading effects in Hadamard transform imaging and spectroscopy systems [J].
Hanley, QS .