Enumeration of peripheral lymphocyte subsets using 6 vs. 4 color staining: A clinical evaluation of a new flowcytometer

被引:16
作者
Lambert, C [1 ]
Cristina, I [1 ]
Christian, G [1 ]
机构
[1] Univ Hosp St Etienne, Immunol Lab, St Etienne, France
关键词
T cell subsets; 6 color flowcytometry; quality assurance; routine analysis; immune monitoring;
D O I
10.1002/cyto.b.20072
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Technological advances in instruments allow the evaluation of many lymphocyte subsets in one step. The :aim of this study was to evaluate the new FACSCanto flowcytometer in routine conditions, using a 6 color combination, single platform, whole blood, lysis, no wash protocol. Methods: Two systems were simultaneously compared on 67 blood samples and external quality controls, :sing CD3,CD4, COB, CD19, CD16/56, and CD45 in one tube TRUCOUNT beads(TM) (BD Biosciences) or two tubes (TetraChrome(TM) and Flowcount(TM), Beckman-Counter and DakoCytomation). Results: The day-to-day instrument detection but automatic compensations were stable. Manual compensation settings were satisfactory using available facilities. Commercial and UK NEQAS quality control results were acceptable. The intra-experiment reproducibility was good (coefficient of variation (CV) <3%) but highly operator-dependant (CD4(+) T cell count CVs from 1.2 to 9.7, six operators). Storage of samples was :acceptable, but storage of stained samples altered absolute count reliability. Serial dilutions show a good count accuracy. The FACScanto T subsets and B cell data were highly correlated with our reference values (r(2) > 0.87) and absolutes values were very close (slopes > 0.89). The gating strategy, fluorochrome choice, and compensation setting are discussed. A few improvements are expected (sample loader, data management, auto-gating, acquisition parameters, sample mixing, absolute values calculation, etc). In conclusions, despite its complexity, 6 color staining is a reliable, stable, and highly informative technique for lymphocyte subset monitoring but remains to be optimized. (C) 2005 International Society for Analytical Cytology.
引用
收藏
页码:29 / 38
页数:10
相关论文
共 34 条
[1]  
Barnett D, 1996, CYTOMETRY, V26, P216, DOI 10.1002/(SICI)1097-0320(19960915)26:3<216::AID-CYTO6>3.0.CO
[2]  
2-C
[3]  
Baudouin F, 1999, CYTOMETRY, V38, P127
[4]  
Bellido M, 1998, HAEMATOLOGICA, V83, P681
[5]   Evaluation of stabilized blood cell products as candidate preparations for quality assessment programs for CD4 T-cell counting [J].
Bergeron, M ;
Shafaie, A ;
Ding, T ;
Phaneuf, S ;
Soucy, N ;
Mandy, F ;
Bradley, J ;
Fahey, J .
CYTOMETRY, 2002, 50 (02) :86-91
[6]   The eosinophilias, including the idiopathic hypereosinophilic syndrome [J].
Brito-Babapulle, F .
BRITISH JOURNAL OF HAEMATOLOGY, 2003, 121 (02) :203-223
[7]   γδ T cells:: Functional plasticity and heterogeneity [J].
Carding, SR ;
Egan, PJ .
NATURE REVIEWS IMMUNOLOGY, 2002, 2 (05) :336-345
[8]  
*CDCP, 1997, MMWR-MORBID MORTAL W, V46, P1
[9]   MULTICOLOR FLOW CYTOMETRIC ANALYSIS OF THE CD45 ANTIGEN PROVIDES IMPROVED LYMPHOID-CELL DISCRIMINATION IN BONE-MARROW AND TISSUE BIOPSIES [J].
FESTIN, R ;
BJORKLAND, A ;
TOTTERMAN, TH .
JOURNAL OF IMMUNOLOGICAL METHODS, 1994, 177 (1-2) :215-224
[10]   Classical and nonclassical class I major histocompatibility complex molecules exhibit subtle conformational differences that affect binding to CD8αα [J].
Gao, GF ;
Willcox, BE ;
Wyer, JR ;
Boulter, JM ;
O'Callaghan, CA ;
Maenaka, K ;
Stuart, DI ;
Jones, EY ;
Van Der Merwe, PA ;
Bell, JI ;
Jakobsen, BK .
JOURNAL OF BIOLOGICAL CHEMISTRY, 2000, 275 (20) :15232-15238