Preliminary evaluation of the new hematology analyzer COULTER® GEN•STM in a university hospital

被引:14
作者
Picard, F [1 ]
Gicquel, C [1 ]
Marnet, L [1 ]
Guesnu, M [1 ]
Levy, JP [1 ]
机构
[1] CHU Cochin Port Royal, Hematol Lab, F-75679 Paris 14, France
关键词
COULTER (R) GEN center dot STM; evaluation; complete blood count; differential count;
D O I
10.1515/CCLM.1999.106
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 [基础医学];
摘要
The COULTER(R) GEN.S(TM) system (COULTER(R) Corp, Miami, USA) is an automated hematology instrument that is designed to provide a complete hematological profile including white blood cells (WBC), complete blood count (CBC) differential count (diff) and the reticulocyte parameters. It was evaluated in our laboratory over a one month period. A preliminary study was performed using the GEN.S(TM) software revision 1D. The evaluation had two purposes: 1) evaluation of the GEN.S(TM) specifications; 2) comparison of its analytical performance with the hematology analyzer currently used in our laboratory. The first part of the evaluation showed that the COULTER(R) GEN.S(TM) is reproducible, has linearity beyond the specifications given by the manufacturer and produces stable results up to 48 hours after blood collection. The evaluation of analytical performance included: 1) a comparison between the GEN.S(TM) CBC and diff numerical results and the COULTER(R) STKS(TM) (COULTER(R) Corp, Miami, USA). These comparisons showed that the results given by the two methods are-similar and suggested that the COULTER(R) GEN.S(TM) could replace the current hematology instrument in use in our laboratory; 2) a performance analysis, to measure the system's ability to detect morphologic abnormalities; when compared to the reference blood smear examination. This part of the evaluation was performed on both normal and abnormal samples. The results of this analysis showed that the sensitivity of the GEN.S(TM) was excellent, especially regarding blast cells, immature granulocytes, nucleated red blood cells (NRBCs) and platelet clumps when using the complete suspect flagging system of the instrument. In addition, the use of the review criteria in our laboratory allowed to detect all hematological diseases. The overall false negative rate was 0.9 %. Thus we consider that the COULTER(R) GEN.S(TM) system is suited for use in medium-to large hospital laboratories which perform more than 100 CBC/day. Overall, the instrument had excellent performance, with a throughput of more than 100 samples/h. It's user friendly workstation has complete patient data management, which is in compliance with good laboratory practices in France.
引用
收藏
页码:681 / 686
页数:6
相关论文
共 18 条
[1]
EVALUATION OF 4 AUTOMATED HEMATOLOGY ANALYZERS - A COMPARATIVE-STUDY OF DIFFERENTIAL COUNTS (IMPRECISION AND INACCURACY) [J].
BUTTARELLO, M ;
GADOTTI, M ;
LORENZ, C ;
TOFFALORI, E ;
CESCHINI, N ;
VALENTINI, A ;
RIZZOTTI, P .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1992, 97 (03) :345-352
[2]
ENGLAND AP, 1991, MED TECHNOL, V5, P1
[3]
FIALON P, 1990, PATHOL BIOL, V38, P743
[4]
FLANDRIN G, 1993, ANN BIOL CLIN, V50, P787
[5]
FOURNIER M, 1994, CLIN LAB HAEMATOL, V16, P33
[6]
MONOCYTE COUNTING - DISCREPANCIES IN RESULTS OBTAINED WITH DIFFERENT AUTOMATED INSTRUMENTS [J].
GOOSSENS, W ;
VANHOVE, L ;
VERWILGHEN, RL .
JOURNAL OF CLINICAL PATHOLOGY, 1991, 44 (03) :224-227
[7]
KESSLER C, 1997, LAB HEMATOL, V3, P41
[8]
Kessler C., 1997, LAB HEMATOL, V3, P32
[9]
LACOMBE C, 1994, NOUV REV FR HEMATOL, V36, P343
[10]
*NAT COMM CLIN LAB, 1992, H20A NCCLS