An electrical knee lock system for functional electrical stimulation

被引:11
作者
Kagaya, H
Shimada, Y
Sato, K
Sato, M
Iizuka, K
Obinata, G
机构
[1] BIOTEC LTD,AKITA,JAPAN
[2] AKITA UNIV,COLL MIN,DEPT MECH ENGN,AKITA,JAPAN
来源
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION | 1996年 / 77卷 / 09期
关键词
D O I
10.1016/S0003-9993(96)90272-5
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
An electrical knee lock system that can be combined with functional electrical stimulation was designed for paraplegic patients. This knee system unlocks the knee electrically and allows knee flexion during the swing phase of the gait. When the knee is extended by electrical stimulation of the knee extensors, the knee is automatically locked by the weight of the locking bar, and the stimulation of the knee extensors is stopped. Since the knee extensors are stimulated for only a short period, muscle fatigue of the knee extensors seldom occurs. We applied this system to a T8 completely paraplegic patient. Standing-up, standing, walking, and sitting-down motions were all restored by our hybrid system. No electrical stimulation was necessary during standing, and the knee extensors were stimulated during only a small percentage of the 1-gait cycle. (C) 1996 by the American Congress of Rehabilitation Medicine and the American Academy of Physical Medicine and Rehabilitation
引用
收藏
页码:870 / 873
页数:4
相关论文
共 13 条
[1]  
ANDREWS BJ, 1989, AUTOMEDICA, V11, P175
[2]  
HIROKAWA S, 1990, ARCH PHYS MED REHAB, V71, P687
[3]   RESTORATION AND ANALYSIS OF STANDING-UP IN COMPLETE PARAPLEGIA UTILIZING FUNCTIONAL ELECTRICAL-STIMULATION [J].
KAGAYA, H ;
SHIMADA, Y ;
EBATA, K ;
SATO, M ;
SATO, K ;
YUKAWA, T ;
OBINATA, G .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 1995, 76 (09) :876-881
[4]   Changes in muscle force following therapeutic electrical stimulation in patients with complete paraplegia [J].
Kagaya, H ;
Shimada, Y ;
Sato, K ;
Sato, M .
PARAPLEGIA, 1996, 34 (01) :24-29
[5]   REPORT ON A CONFERENCE ON MOTOR PROSTHESES FOR WORKPLACE MOBILITY OF PARAPLEGIC PATIENTS IN NORTH-AMERICA [J].
KANTOR, C ;
ANDREWS, BJ ;
MARSOLAIS, EB ;
SOLOMONOW, M ;
LEW, RD ;
RAGNARSSON, KT .
PARAPLEGIA, 1993, 31 (07) :439-456
[6]  
MARSOLAIS EB, 1988, ARCH PHYS MED REHAB, V69, P243
[7]   FUNCTIONAL ELECTRICAL-STIMULATION FOR WALKING IN PARAPLEGIA [J].
MARSOLAIS, EB ;
KOBETIC, R .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1987, 69A (05) :728-733
[8]   EXPERIMENTAL-STUDY OF A SENSOR-CONTROLLED EXTERNAL KNEE LOCKING SYSTEM [J].
MCGHEE, RB ;
TOMOVIC, R ;
YANG, PY ;
MACLEAN, IC .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1978, 25 (02) :195-199
[9]  
POPOVIC D, 1987, P INT S EXT CONTR HU, P95
[10]  