Determination of chromium in urine by electrothermal atomic absorption spectrometry using different chemical modifiers

被引:33
作者
Burguera, JL
Burguera, M
Rondon, C
Rodríguez, L
Carrero, P
de Peña, YP
Burguera, E
机构
[1] Univ Los Andes, Fac Sci, Venezuelan Andean Inst Chem Res, IVAIQUIM, Merida 5101A, Venezuela
[2] Univ Los Andes, Fac Odontol, Dept Prevent & Social Odontol, Merida 5101A, Venezuela
关键词
D O I
10.1039/a808239a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The effects of the chemical modifiers Eu, Mg(NO3)(2), Pd, Eu-Pd and Ni and the use of longitudinally (with deuterium lamp background correction) and transversally (with Zeeman effect background correction) heated atomizers on the determination of chromium in urine samples were studied. When longitudinally heated atomizers were used, the background absorption was too high for the addition of most species, such as Mg(NO3)(2), Pd, Eu-Pd and Ni, rendering its use unsuitable for the determination of chromium in urine. The addition of Eu led to lower background signals. However, when longitudinally heated atomizers were used larger background absorption signals were observed on addition of a modifier. In contrast, with transversally heated atomizers the background absorption signals were low and were not influenced by the addition of a modifier, except when Mg(NO3)(2) was added. In this case, the atomization signal shapes and the sensitivity were improved. The characteristic masses (m(0)) and limits of detection (3 sigma) were 2.2 and 3.3 pg and 0.02 and 0.04 mu g Cr l(-1) for the longitudinally (using Eu as modifier) and transversally [using Mg(NO3)(2)] heated atomizers, respectively. Recovery studies and analysis of standard reference materials certified for chromium were performed to asses the accuracy. The results for the determination of chromium in real samples with both background correction procedures agreed well with a precision between 0.8 and 2.5%. Both procedures can be recommended and the choice will depend on instrument availability.
引用
收藏
页码:821 / 825
页数:5
相关论文
共 29 条
[1]   Determination of nickel in saliva by electrothermal atomic absorption spectrometry using various chemical modifiers with Zeeman-effect background correction [J].
Burguera, E ;
De Briceno, AS ;
Rondon, CE ;
Burguera, JL ;
Burguera, M ;
Carrero, P .
JOURNAL OF TRACE ELEMENTS IN MEDICINE AND BIOLOGY, 1998, 12 (02) :115-120
[2]  
Burguera JL, 1997, ATOM SPECTROSC, V18, P109
[3]  
Burguera M, 1998, METAL IONS BIOL MED, V5, P13
[4]  
BURGUERA M, 1999, SPECTROCHIM ACTA B
[5]   PHYSIOLOGICAL CHROMIUM DETERMINATION IN SERUM BY ZEEMAN GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY - A SERIOUS CHALLENGE [J].
CHAPPUIS, P ;
POUPON, J ;
DESCHAMPS, JF ;
GUILLAUSSEAU, PJ ;
ROUSSELET, F .
BIOLOGICAL TRACE ELEMENT RESEARCH, 1992, 32 :85-91
[6]   DETERMINATION OF CHROMIUM IN BIOLOGICAL-FLUIDS BY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY USING WALL, PLATFORM AND PROBE ATOMIZATION FROM DIFFERENT GRAPHITE SURFACES [J].
CIMADEVILLA, EAC ;
WROBEL, K ;
GAYON, JMM ;
SANZMEDEL, A .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (02) :117-123
[7]  
GENE O, 1981, SPECTROCHIM ACTA B, V36, P163
[8]  
GONG BL, 1990, ATOM SPECTROSC, V11, P229
[9]  
Goyer R.A., 1986, CASARETT DOULLS TOXI
[10]   DETERMINATION OF TOTAL CHROMIUM IN WHOLE-BLOOD, BLOOD COMPONENTS, BONE, AND URINE BY FAST FURNACE PROGRAM ELECTROTHERMAL ATOMIZATION AAS AND USING NEITHER ANALYTE ISOFORMATION NOR BACKGROUND CORRECTION [J].
GRANADILLO, VA ;
DEMACHADO, LP ;
ROMERO, RA .
ANALYTICAL CHEMISTRY, 1994, 66 (21) :3624-3631