Integrated electroanatomic mapping with three-dimensional computed tomographic images for real-time guided ablations

被引:186
作者
Dong, J
Calkins, H
Solomon, SB
Lai, SH
Dalal, D
Lardo, A
Brem, E
Preiss, A
Berger, RD
Halperin, H
Dickfeld, T
机构
[1] Univ Maryland, Baltimore, MD 21201 USA
[2] Johns Hopkins Univ, Sch Med, Baltimore, MD USA
[3] Bloomberg Sch Publ Hlth, Baltimore, MD USA
[4] Biosense Webster Inc, Haifa, Israel
关键词
catheter ablation; electrophysiology; imaging; mapping;
D O I
10.1161/CIRCULATIONAHA.105.565200
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background - New ablation strategies for atrial fibrillation or nonidiopathic ventricular tachycardia are increasingly based on anatomic consideration and require the placement of ablation lesions at the correct anatomic locations. This study sought to evaluate the accuracy of the first clinically available image integration system for catheter ablation on 3-dimensional (3D) computed tomography (CT) images in real time. Methods and Results - After midline sternotomy, 2.3-mm CT fiducial markers were attached to the epicardial surface of each cardiac chamber in 9 mongrel dogs. Detailed 3D cardiac anatomy was reconstructed from contrast-enhanced, high-resolution CT images and registered to the electroanatomic maps of each cardiac chamber. To assess accuracy, targeted ablations were performed at each of the fiducial markers guided only by the reconstructed 3D images. At autopsy, the position error was 1.9 +/- 0.9 mm for the right atrium, 2.7 +/- 1.2 mm for the right ventricle, 1.8 +/- 1.0 mm for the left atrium, and 2.3 +/- 1.1 mm for the left ventricle. To evaluate the system's guidance of more complex clinical ablation strategies, ablations of the cavotricuspid isthmus ( n = 4), fossa ovalis ( n = 4), and pulmonary veins ( n = 6) were performed, which resulted in position errors of 1.8 +/- 1.5, 2.2 +/- 1.3, and 2.1 +/- 1.2 mm, respectively. Retrospective analysis revealed that a combination of landmark registration and the target chamber surface registration resulted in < 3 mm accuracy in all 4 cardiac chambers. Conclusions - Image integration with high-resolution 3D CT allows accurate placement of anatomically guided ablation lesions and can facilitate complex ablation strategies. This may provide significant advantages for anatomically based procedures such as ablation of atrial fibrillation and nonidiopathic ventricular tachycardia.
引用
收藏
页码:186 / 194
页数:9
相关论文
共 21 条
  • [1] Nonfluoroscopic, in vivo navigation and mapping technology
    BenHaim, SA
    Osadchy, D
    Schuster, I
    Gepstein, L
    Hayam, G
    Josephson, ME
    [J]. NATURE MEDICINE, 1996, 2 (12) : 1393 - 1395
  • [2] De Groot NM, 2000, J CARDIOVASC ELECTR, V11, P1183
  • [3] Stereotactic catheter navigation using magnetic resonance image integration in the human heart
    Dickfeld, T
    Calkins, H
    Bradley, D
    Solomon, SB
    [J]. HEART RHYTHM, 2005, 2 (04) : 413 - 415
  • [4] Anatomic stereotactic catheter ablation on three-dimensional magnetic resonance images in real time
    Dickfeld, T
    Calkins, H
    Zviman, M
    Kato, R
    Meininger, G
    Lickfett, L
    Berger, R
    Halperin, H
    Solomon, SB
    [J]. CIRCULATION, 2003, 108 (19) : 2407 - 2413
  • [5] Stereotactic magnetic resonance guidance for anatomically targeted ablations of the fossa ovalis and the left atrium
    Dickfeld, T
    Calkins, H
    Zviman, M
    Meininger, G
    Lickfett, L
    Roguin, A
    Lardo, AC
    Berger, R
    Halperin, H
    Solomon, SB
    [J]. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2004, 11 (02) : 105 - 115
  • [6] Gepstein L, 1997, CIRCULATION, V95, P1611
  • [7] Hochberg Y., 1987, Multiple comparison procedures
  • [8] Pulmonary vein anatomy in patients undergoing catheter ablation of atrial fibrillation - Lessons learned by use of magnetic resonance imaging
    Kato, R
    Lickfett, L
    Meininger, G
    Dickfeld, T
    Wu, R
    Juang, G
    Angkeow, P
    LaCorte, J
    Bluemke, D
    Berger, R
    Halperin, HR
    Calkins, H
    [J]. CIRCULATION, 2003, 107 (15) : 2004 - 2010
  • [9] Characterization of a new pulmonary vein variant using magnetic resonance angiography: Incidence, imaging, and interventional implications of the "right top pulmonary vein"
    Lickfett, L
    Kato, R
    Tandri, H
    Jayam, V
    Vasamreddy, CR
    Dickfeld, T
    Lewalter, T
    Luderitz, B
    Berger, R
    Halperin, H
    Calkins, H
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2004, 15 (05) : 538 - 543
  • [10] Linear ablation lesions for control of unmappable ventricular tachycardia in patients with ischemic and nonischemic cardiomyopathy
    Marchlinski, FE
    Callans, DJ
    Gottlieb, CD
    Zado, E
    [J]. CIRCULATION, 2000, 101 (11) : 1288 - 1296