Retrieved glenoid components - A classification system for surface damage analysis

被引:50
作者
Gunther, SB [1 ]
Graham, J [1 ]
Norris, TR [1 ]
Ries, MD [1 ]
Pruitt, L [1 ]
机构
[1] Univ Calif San Francisco, Dept Orthopaed Surg, San Francisco, CA 94143 USA
关键词
polyethylene wear; shoulder arthroplasty; glenoid failure; surface damage classification;
D O I
10.1054/arth.2002.27671
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
There have been many reports describing modes of damage in retrieved total hip and total knee arthroplasty components. The most common mechanism in total hip arthroplasties has been shown to be surface wear. Fatigue failure shown as pitting and delamination are observed more often in total knee components. There has been no previous analysis of retrieved polyethylene glenoid components. This study evaluate the wear mechanisms contributing to failure of total shoulder glenoid components. Polyethylene glenoid components from 10 consecutive total shoulder arthroplasties have been retrieved and analyzed. Wear mechanisms were analyzed under low-power magnification, and a classification system was designed for total shoulder arthroplasties. This classification system is an adaptation of previous models of hip and knee surface damage. The severity of each damage mode was graded in 4 separate quadrants. The most prevalent damage modes were abrasion, pitting, and delamination. These data show a combination of abrasive wear and fatigue in retrieved total shoulder specimens. Surface wear and subsurface fatigue failure mechanisms both contribute to glenoid implant failure.
引用
收藏
页码:95 / 100
页数:6
相关论文
共 19 条
[1]
Baker DA, 1999, J BIOMED MATER RES, V46, P573, DOI 10.1002/(SICI)1097-4636(19990915)46:4<573::AID-JBM16>3.0.CO
[2]
2-A
[3]
THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[4]
Wear in retrieved condylar knee arthroplasties - A comparison of wear in different designs of 280 retrieved condylar knee prostheses [J].
Blunn, GW ;
Joshi, AB ;
Minns, RJ ;
Lidgren, L ;
Lilley, P ;
Ryd, L ;
Engelbrecht, E ;
Walker, PS .
JOURNAL OF ARTHROPLASTY, 1997, 12 (03) :281-290
[5]
COLLIER JP, 1991, CLIN ORTHOP RELAT R, P232
[6]
A QUANTITATIVE TECHNIQUE FOR REPORTING SURFACE DEGRADATION PATTERNS OF UHMWPE COMPONENTS OF RETRIEVED TOTAL KNEE REPLACEMENTS [J].
CORNWALL, GB ;
BRYANT, JT ;
HANSSON, CM ;
RUDAN, J ;
KENNEDY, LA ;
COOKE, TDV .
JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (01) :9-18
[7]
GOODMAN SB, 1993, SEMIN ARTHROPLASTY, V4, P205
[8]
LIGHT-MICROSCOPIC IDENTIFICATION OF SUBMICRON POLYETHYLENE WEAR DEBRIS [J].
GUTTMANN, D ;
SCHMALZRIED, TP ;
JASTY, M ;
HARRIS, WH .
JOURNAL OF APPLIED BIOMATERIALS, 1993, 4 (04) :303-307
[9]
RETRIEVAL ANALYSIS OF TOTAL KNEE PROSTHESES - A METHOD AND ITS APPLICATION TO 48 TOTAL CONDYLAR PROSTHESES [J].
HOOD, RW ;
WRIGHT, TM ;
BURSTEIN, AH .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1983, 17 (05) :829-842
[10]
STUDIES OF THE MECHANISM BY WHICH THE MECHANICAL FAILURE OF POLYMETHYLMETHACRYLATE LEADS TO BONE-RESORPTION [J].
HOROWITZ, SM ;
DOTY, SB ;
LANE, JM ;
BURSTEIN, AH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1993, 75A (06) :802-813