Excimer laser treatment of NiTi shape memory alloy biomaterials

被引:87
作者
Villermaux, F
Tabrizian, M
Yahia, LH
Meunier, M
Piron, DL
机构
[1] ECOLE POLYTECH, COUCHES MINCES GRP, DEPT ENGN PHYS, MONTREAL, PQ H3C 3A7, CANADA
[2] ECOLE POLYTECH, DEPT MET, MONTREAL, PQ H3C 3A7, CANADA
关键词
NiTi shape memory alloy; laser surface treatment; corrosion resistance; biomaterials;
D O I
10.1016/S0169-4332(96)00619-8
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
NiTi Shape Memory Alloys (SMA) are potential biomaterial candidates. However, due to its possible corrosion in physiological solution, dissolution of toxic Ni might be happening, rendering this material nonbiocompatible. We have used excimer laser surface treatment to improve corrosion resistance of NiTi SMA plates. Potentiodynamic tests in physiological Hank's solution show that the laser treatment performed in air improved all corrosion parameters. The surface is homogenized and a Scanning Electron Microscopy (SEM) observation indicates a decrease of corrosion pit size and numbers. Laser treatment improvement resistance is explained by a combination of the homogenization of the surface by melting, the hardening due to N incorporation and the thickening of the oxide layer.
引用
收藏
页码:62 / 66
页数:5
相关论文
共 19 条
[1]  
ABIKIDO Y, 1994, P 1 INT C SHAP MEM S
[2]  
BERGERON A, 1994, TRAITEMENT SURFACES
[3]   MODIFICATION OF SURFACE-FILMS ON METALLIC SUBSTRATES BY EXCIMER LASER IRRADIATION [J].
BERGMANN, HW ;
SCHUBERT, E ;
SCHMATJKO, KJ ;
DEMBOWSKI, J .
THIN SOLID FILMS, 1989, 174 :33-38
[4]  
Castleman L.S., 1981, BIOCOMPATIBILITY CLI, V1, P129
[5]  
DUTTA RS, 1993, BRIT CORROS J, V28, P217
[6]  
EDIE JW, 1981, ANGLE ORTHOD, V51, P319
[7]   EFFECT OF N+ IMPLANTATION ON THE SHAPE-MEMORY BEHAVIOR AND CORROSION-RESISTANCE OF AN EQUIATOMIC NI TI ALLOY [J].
GREEN, SM ;
GRANT, DM ;
WOOD, JV ;
JOHANSON, A ;
JOHNSON, E ;
SARHOLTKRISTENSEN, L .
JOURNAL OF MATERIALS SCIENCE LETTERS, 1993, 12 (09) :618-619
[8]  
KIMURA H, 1987, DENT MAT APPARATUS, V6, P73
[9]  
LOMBARDI S, 1994, P 1 INT C SHAP MEM S
[10]   INVIVO MEASUREMENT OF ANODIC POLARIZATION OF ORTHOPEDIC IMPLANT ALLOYS - COMPARATIVE-STUDY OF INVIVO AND INVITRO EXPERIMENTS [J].
NAKAYAMA, Y ;
YAMAMURO, T ;
KOTOURA, Y ;
OKA, M .
BIOMATERIALS, 1989, 10 (06) :420-424