Use of the BD PHOENIX automated identification and susceptibility testing positive blood cultures in a microbiology system for direct of gram-negative rods from three-phase trial

被引:74
作者
Funke, G [1 ]
Funke-Kissling, P [1 ]
机构
[1] Gartner & Colleagues Labs, Dept Med Microbiol & Hyg, D-88250 Weingarten, Germany
关键词
D O I
10.1128/JCM.42.4.1466-1470.2004
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
The present study describes the use of the automated BACTEC 9240 blood culture system, the Serum Separator Tube (SST), and the BD PHOENIX Automated Microbiology System in combination for the direct identification and antimicrobial susceptibility testing (AST) of gram-negative rods (GNRs) from positive blood cultures (BCs) without subculture. The study was conducted in three phases: (i) the recovery yield of Escherichia coli ATCC 25922 was determined with the SST between 0 and 8 h after spiked BC bottles turned positive; (ii) the identifications and susceptibility testing results obtained with the PHOENIX system for nine American Type Culture Collection strains of GNRs processed by the SST procedure and for colonies from agar medium were compared; and (iii) the procedure with the BACTEC system, SSTs, and the PHOENIX system was applied to positive cultures of blood from 309 patients during a 3-month period. The SST procedure with E. coli yielded sufficient numbers of cells to perform direct inoculation at any time between 0 and 8 h after a BC bottle turned positive. By using the identities obtained from pure cultures with the PHOENIX system and other biochemical identification systems as reference methods, the agreement between the reference methods and the PHOENIX system tested directly by using cultures of blood from patients was 92.9%. The 7.1% discrepant results were due to 6.5% incorrect identifications with the PHOENIX system with BC samples and 0.6% incorrect identifications with the PHOENIX system with samples from agar cultures. By AST the overall categorical accuracy was 99.0%, with 0.1% very major errors, 0.1% major errors, and 0.8% minor errors. In conclusion, use of the combination of the BACTEC system, SSTs, and the PHOENIX system has the potential to allow the agar isolation step to be skipped and the procedures for rapid direct identification and susceptibility testing of GNRs from positive BCs to be improved both in hospital-based and in central non-hospital-based laboratories.
引用
收藏
页码:1466 / 1470
页数:5
相关论文
共 13 条
[1]   Clinical and financial benefits of rapid bacterial identification and antimicrobial susceptibility testing [J].
Barenfanger, J ;
Drake, C ;
Kacich, G .
JOURNAL OF CLINICAL MICROBIOLOGY, 1999, 37 (05) :1415-1418
[2]   CLINICAL IMPACT OF RAPID IN-VITRO SUSCEPTIBILITY TESTING AND BACTERIAL IDENTIFICATION [J].
DOERN, GV ;
VAUTOUR, R ;
GAUDET, M ;
LEVY, B .
JOURNAL OF CLINICAL MICROBIOLOGY, 1994, 32 (07) :1757-1762
[3]   Evaluation of wider system for direct identification and antimicrobial susceptibility testing of gram-negative bacilli from positive blood culture bottles [J].
Fontanals, D ;
Salceda, F ;
Hernández, J ;
Sanfeliu, I ;
Torra, M .
EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2002, 21 (09) :693-695
[4]   Direct identification and susceptibility testing of enteric bacilli from positive blood cultures using VITEK (GNI plus /GNS-GA) [J].
Hansen, DS ;
Jensen, AG ;
Norskov-Lauritsen, N ;
Skov, R ;
Bruun, B .
CLINICAL MICROBIOLOGY AND INFECTION, 2002, 8 (01) :38-44
[5]   Evaluation of the VITEK 2 system for rapid direct identification and susceptibility testing of gram-negative bacilli from positive blood cultures [J].
Ling, TKW ;
Liu, ZK ;
Cheng, AFB .
JOURNAL OF CLINICAL MICROBIOLOGY, 2003, 41 (10) :4705-4707
[6]   EVALUATING BIOCHEMICAL-IDENTIFICATION SYSTEMS [J].
MILLER, JM .
JOURNAL OF CLINICAL MICROBIOLOGY, 1991, 29 (08) :1559-1561
[7]  
Murray PR, 2003, MANUAL CLIN MICROBIO
[8]  
National Committee for Clinical Laboratory Standards, 2003, PERF STAND ANT DISK
[9]   Accuracy of the vitek system for antimicrobial susceptibility testing Enterobacteriaceae bloodstream infection isolates: Use of ''direct'' inoculation from Bactec 9240 blood culture bottles [J].
Putnam, LR ;
Howard, WJ ;
Pfaller, MA ;
Koontz, FP ;
Jones, RN .
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 1997, 28 (02) :101-104
[10]   Evaluation of the BD Phoenix automated identification and susceptibility testing system in clinical microbiology laboratory practice [J].
Stefaniuk, E ;
Baraniak, A ;
Gniadkowski, M ;
Hryniewicz, W .
EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2003, 22 (08) :479-485