The reliability and validity of two Ambulatory Monitoring actigraphs

被引:25
作者
Tryon, WW [1 ]
机构
[1] Fordham Univ, Dept Psychol, Bronx, NY 10458 USA
关键词
D O I
10.3758/BF03192719
中图分类号
B841 [心理学研究方法];
学科分类号
040201 ;
摘要
Evidence for the reliability and validity of two models of Ambulatory Monitoring, Inc. actigraphs was obtained by testing four instruments of each kind 10 times each on a precision pendulum. Correlation and coefficient of variation methods were used to analyze the data. Reliability coefficients of .98 were obtained for both models. Coefficient of variation methods yielded reliability coefficients of 92% for the MotionLogger model and 97% for the BuzzBee model. Validity coefficients of. 99 were obtained for both models. However, MotionLogger means were found to differ by up to 36% from one instrument to another, whereas BuzzBee means differed by 10% at most. Issues of standardization and measurement units were discussed.
引用
收藏
页码:492 / 497
页数:6
相关论文
共 30 条
[1]  
Anastasi A., 1997, Psychological testing
[2]  
[Anonymous], J INTERAMER MED
[3]  
BOUTEN CV, 1994, MED SCI SPORT EXER, V26, P1516
[4]   INSTRUMENTED MOTOR-ACTIVITY MEASUREMENT OF THE YOUNG INFANT IN THE HOME - VALIDITY AND RELIABILITY [J].
EATON, WO ;
MCKEEN, NA ;
LAM, CS .
INFANT BEHAVIOR & DEVELOPMENT, 1988, 11 (03) :375-378
[5]   Calibration of the Computer Science and Applications, Inc. accelerometer [J].
Freedson, PS ;
Melanson, E ;
Sirard, J .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 1998, 30 (05) :777-781
[6]   ELECTRONIC MOTION SENSORS AND HEART-RATE AS MEASURES OF PHYSICAL-ACTIVITY IN CHILDREN [J].
FREEDSON, PS .
JOURNAL OF SCHOOL HEALTH, 1991, 61 (05) :220-223
[7]  
Ghiselli E.E., 1981, Measurement theory for the behavioral sciences
[8]  
Ghiselli E. E., 1964, Theory of psychological measurement
[9]  
Heil D. P., 2003, Medicine & Science in Sports & Exercise, V35, pS285, DOI 10.1097/00005768-200305001-01586