Solid-phase microextraction for the analysis of human breath

被引:238
作者
Grote, C [1 ]
Pawliszyn, J [1 ]
机构
[1] UNIV WATERLOO, DEPT CHEM, GUELPH WATERLOO CTR GRAD WORK CHEM, WATERLOO, ON N2L 3G1, CANADA
关键词
D O I
10.1021/ac960749l
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Solid-phase microextraction (SPRaE) has been applied to the quantitative determination of ethanol, acetone, and isoprene in human breath. The method involves extraction and preconcentration with a fused silica fiber coated with a polymeric stationary phase, desorption at 200 degrees C, and assay by gas chromatography/mass spectrometry, Three different fiber coatings have been evaluated with regard to sensitivity, linear range, precision, and detection limits. Typical RSD values in the range 2%-6% could be obtained, depending on the fiber coating and the compound investigated. The calibration curves for the compounds are reproducible and linear over the concentration ranges found in human breath samples. The method is capable of detecting concentrations of acetone and isoprene reported for healthy subjects. The influence of temperature and humidity on the extraction process has been studied in detail, A linear relationship between log K versus 1/T allows the calibration of the method for any given temperature, The device is portable, economical, and easy to use in patient sampling.
引用
收藏
页码:587 / 596
页数:10