Poor reproducibility of in-source collisional atmospheric pressure ionization mass spectra of toxicologically relevant drugs

被引:50
作者
Bogusz, MJ [1 ]
Maier, RD
Krüger, KD
Webb, KS
Romeril, J
Miller, ML
机构
[1] Aachen Univ Technol, Inst Forens Med, D-52057 Aachen, Germany
[2] Lab Govt Chemist, Teddington TW11 0LY, Middx, England
[3] Fed Bur Invest Acad, Forens Sci Res & Teaching Ctr, Quantico, VA 22135 USA
关键词
inter-laboratory study; mass spectrometry; morphine-6-glucuronide; monoacetylmorphine; codeine; lysergic acid diethylamide; methylenedioxymethamphetamine; amphetamines;
D O I
10.1016/S0021-9673(99)00312-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The purpose of the study was to examine the intra- and interlaboratory reproducibility of mass spectra obtained with liquid chromatography-atmospheric pressure ionization mass spectrometry (LC-API-MS) both in electrospray (ESI) and atmospheric pressure chemical ionization (APCI) modes. Toxicologically relevant drugs of different polarity were selected as test substances: morphine-6-glucuronide, 6-monoacetylmorphine, codeine, lysergic acid diethylamide, methylenedioxymethamphetamine. The study was performed in two laboratories using identical instruments and in one using a slightly different instrument. Basic instrument settings and mobile phase were identical in all laboratories. Mass spectra of drugs were taken at four collision energy voltages and using mobile phase of different composition (four concentration levels of acetonitrile and of ammonium formate buffer). The experiments demonstrated that mass spectra of given drugs, obtained in identical conditions with identical instruments, may show very different degrees of fragmentation. Mass spectra obtained with different instruments differed profoundly not only in the degree of fragmentation, but also different fragments and adducts were observed. Short-term intralaboratory reproducibility of mass spectra was satisfactory. On the other hand, the long-term experiments showed different degrees of fragmentation of APCI-generated mass spectra at nominally identical fragmentation energy. The changes in the composition of the mobile phase (concentration of organic modifier or buffer molarity) did not affect the reproducibility of fragmentation to any relevant degree. The study showed that the interlaboratory exchange and use of mass spectrum library, generated by single-quadrupole LC-API-MS instruments, is hardly feasible at the moment, even under very carefully standardized conditions. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:409 / 418
页数:10
相关论文
共 19 条
[1]  
ARDREY RE, 1985, PHARM MASS SPECTRA
[2]   Determination of common drugs of abuse in body fluids using one isolation procedure and liquid chromatography atmospheric pressure chemical ionization mass spectrometry [J].
Bogusz, MJ ;
Maier, RD ;
Kruger, KD ;
Kohls, U .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1998, 22 (07) :549-558
[3]   Determination of morphine and its 3- and 6-glucuronides, codeine, codeine-glucuronide and 6-monoacetylmorphine in body fluids by liquid chromatography atmospheric pressure chemical ionization mass spectrometry [J].
Bogusz, MJ ;
Maier, RD ;
Erkens, M ;
Driessen, S .
JOURNAL OF CHROMATOGRAPHY B, 1997, 703 (1-2) :115-127
[4]   Morphine, morphine-3-glucuronide, morphine-6-glucuronide, and 6-monoacetylmorphine determined by means of atmospheric pressure chemical ionization-mass spectrometry liquid chromatography in body fluids of heroin victims [J].
Bogusz, MJ ;
Maier, RD ;
Driessen, S .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1997, 21 (05) :346-355
[5]  
BOGUSZ MJ, IN PRESS J CHROMAT B
[6]   Overview of the applications of liquid chromatography mass spectrometry interfacing systems in food analysis: naturally occurring substances in food [J].
Careri, M ;
Mangia, A ;
Musci, M .
JOURNAL OF CHROMATOGRAPHY A, 1998, 794 (1-2) :263-297
[7]   Applications of liquid chromatography mass spectrometry interfacing systems in food analysis: Pesticide, drug and toxic substance residues [J].
Careri, M ;
Mangia, A ;
Musci, M .
JOURNAL OF CHROMATOGRAPHY A, 1996, 727 (02) :153-184
[8]   BIOMEDICAL AND BIOCHEMICAL APPLICATIONS OF LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
GELPI, E .
JOURNAL OF CHROMATOGRAPHY A, 1995, 703 (1-2) :59-80
[9]   Applications of liquid chromatography mass spectrometry in analytical toxicology: A review [J].
Hoja, H ;
Marquet, P ;
Verneuil, B ;
Lotfi, H ;
Penicaut, B ;
Lachatre, G .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1997, 21 (02) :116-126
[10]   Liquid chromatography mass spectrometry in forensic and clinical toxicology [J].
Maurer, HH .
JOURNAL OF CHROMATOGRAPHY B, 1998, 713 (01) :3-25