Novel bioartificial liver support system: Preclinical evaluation

被引:42
作者
Patzer, JF
Mazariegos, GV
Lopez, R
Molmenti, E
Gerber, D
Riddervold, F
Khanna, A
Yin, WY
Chen, Y
Scott, VL
Aggarwal, S
Kramer, DJ
Wagner, RA
Zhu, Y
Fulmer, ML
Block, GD
Amiot, BP
机构
[1] Univ Pittsburgh, Dept Surg, Pittsburgh, PA 15261 USA
[2] Univ Pittsburgh, Thomas E Starzl Transplantat Inst, Pittsburgh, PA 15261 USA
[3] Univ Pittsburgh, Dept Chem Engn, Pittsburgh, PA 15261 USA
[4] Univ Pittsburgh, McGowan Ctr Artificial Organs, Pittsburgh, PA 15261 USA
[5] Univ Pittsburgh, Dept Anesthesiol, Pittsburgh, PA 15261 USA
[6] Univ Pittsburgh, Cent Anim Facil, Pittsburgh, PA 15261 USA
[7] Univ Monterrey, Dept Surg, Monterrey, NL, Mexico
[8] Univ Oslo, Natl Hosp, Rikshosp, Dept Anesthesiol, N-0027 Oslo 1, Norway
[9] Tzu Chi Gen Hosp, Hualien, Taiwan
[10] Excorp Med Inc, Oakdale, MN 55128 USA
来源
BIOARTIFICIAL ORGANS II: TECHNOLOGY, MEDICINE, AND MATERIALS | 1999年 / 875卷
关键词
D O I
10.1111/j.1749-6632.1999.tb08516.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Preclinical safety and efficacy evaluation of a novel bioartificial liver support system (BLSS) was conducted using a D-galactosamine canine liver failure model, The BLSS houses a suspension of porcine hepatocytes In a hollow fiber cartridge with the hepatocytes on one side of the membrane and whole blood flowing on the other. Porcine hepatocytes harvested by a collagenase digestion technique were infused into the hollow fiber cartridge and incubated for 16 to 24 hours prior to use. Fifteen purpose-bred male hounds, 1-3 years old, 25-30 kg, were administered a lethal dose, 1.5 g/kg, of D-galactosamine. The animals were divided into three treatment groups: (Ib) no BLSS treatment (n = 6); (2b) BLSS treatment starting at 24-26 h post D-galactosamine (n = 5); and (2c) BLSS treatment starting at 16-18 h past D-galactosamine (n = 4). While maintained under isoflurane anesthesia, canine supportive care was guided by electrolyte and invasive physiologic monitoring consisting of arterial pressure, central venous pressure, extradural intracranial pressure (ICP), pulmonary artery pressure, urinary catheter, and end-tidal CO2, All animals were treated until death or death-equivalent (inability to sustain systolic blood pressure > 80 mmHg for 20 minutes despite massive fluid resuscitation and/or dopamine administration), or euthanized at 60 hours, All animals developed evidence of liver failure at 12-24 hours as evidenced by blood pressure lability, elevated ICP, marked hepatocellular enzyme elevation with microscopic massive hepatocyte necrosis and cerebral edema, elevated prothrombin time, and metabolic acidosis. Groups 2b and 2c marginally prolong survival compared with Group Ib (pairwise log rank censored survival time analysis, p = 0.096 and p = 0.064, respectively). Since survival times for Groups 2b and 2c are not significantly different (p = 0.693), the groups were combined for further statistical analysis. Survival times for the combined active treatment Groups 2b and 2c are significantly prolonged versus Group Ib (p = 0.047). These results suggest the novel BLSS reported here can have a significant impact on the course of liver failure in the D-galactosamine canine liver failure model. The BLSS is ready for Phase I safety evaluation in a clinical setting.
引用
收藏
页码:340 / 352
页数:13
相关论文
共 36 条
[1]  
Ash S R, 1993, ASAIO J, V39, pM675
[2]   Push-pull sorbent based pheresis for treatment of acute hepatic failure - The biologic-detoxifier plasma filter system [J].
Ash, SR ;
Blake, DE ;
Carr, DJ ;
Harker, KD .
ASAIO JOURNAL, 1998, 44 (03) :129-139
[3]   Microfabrication of hepatocyte/fibroblast co-cultures: Role of homotypic cell interactions [J].
Bhatia, SN ;
Balis, UJ ;
Yarmush, ML ;
Toner, M .
BIOTECHNOLOGY PROGRESS, 1998, 14 (03) :378-387
[4]  
Bhatia SN, 1997, J BIOMED MATER RES, V34, P189, DOI 10.1002/(SICI)1097-4636(199702)34:2<189::AID-JBM8>3.0.CO
[5]  
2-M
[6]  
BLOCK GD, 1998, UNPUB
[7]   Consideration of potential immunological problems in the application of xenogenic hybrid liver support [J].
Bornemann, R ;
Smith, MD ;
Gerlach, JC .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1996, 19 (11) :655-663
[8]   Importance of the kinetic characterization of liver cell metabolic reactions to the design of hybrid liver support devices [J].
Catapano, G ;
DeBartolo, L .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1996, 19 (11) :670-676
[9]  
Chen SC, 1997, ANN NY ACAD SCI, V831, P350
[10]   Clinical experience with a porcine hepatocyte-based liver support system [J].
Chen, SC ;
Hewitt, WR ;
Watanabe, FD ;
Eguchi, S ;
Kahaku, E ;
Middleton, Y ;
Rozga, J ;
Demetriou, AA .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1996, 19 (11) :664-669