Improved calibrator for the evaluation of batch produced strain gauges in cryogenic environments

被引:4
作者
Huber, WH [1 ]
Hernandez, LM [1 ]
Goldman, AM [1 ]
机构
[1] Univ Minnesota, Sch Phys & Astron, Minneapolis, MN 55455 USA
关键词
D O I
10.1063/1.1149811
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have constructed a simple, improved apparatus for calibrating strain gauges down to liquid helium temperatures to permit an estimate of the repeatability and linearity of commercially available, batch produced, metal foil strain gauges. The apparatus is based upon an original design first proposed by McClintock [R. M. McClintock, Rev. Sci. Instrum. 30, 715 (1959)]. Calibrations performed on a particular batch produced metal foil strain gauge indicate that it behaved ideally at low temperatures. Although the gauge resistance was strongly temperature dependent, the gauge factor was virtually independent of temperature and did not display any detectable nonlinearity or hysteresis. (C) 1999 American Institute of Physics. [S0034-6748(99)04506-2].
引用
收藏
页码:2872 / 2874
页数:3
相关论文
共 9 条
[1]  
ALTMAN HW, 1954, 26427 OH STAT U RES
[2]  
CORRUCCINI RJ, 1961, NBS MONOGR, V29
[3]   STRESS-ANALYSIS DOWN TO LIQUID-HELIUM TEMPERATURE [J].
FERRERO, C .
CRYOGENICS, 1990, 30 (03) :249-254
[4]  
FREYNIK HS, 1976, ADV CRYOGEN ENG, V24, P473
[5]   STRAIN-GAUGES FOR MEASUREMENT OF MAGNETOSTRICTION IN RANGE 4K TO 300K [J].
GREENOUGH, RD ;
UNDERHILL, C .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1976, 9 (06) :451-454
[6]   STRAIN GAUGE CALIBRATION DEVICE FOR EXTREME TEMPERATURES [J].
MCCLINTOCK, RM .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1959, 30 (08) :715-718
[7]  
Pobell F, 1992, Matter and Methods at Low Temperatures
[8]   STRAIN GAGES IN CRYOGENIC ENVIRONMENT [J].
TELINDE, JC .
EXPERIMENTAL MECHANICS, 1970, 10 (09) :394-+
[9]  
Window A. L., 1992, STRAIN GAUGE TECHNOL