Design and performance of the UCLH Mark 1b 64 channel electrical impedance tomography (EIT) system, optimized for imaging brain function

被引:44
作者
Yerworth, RJ [1 ]
Bayford, RH
Cusick, G
Conway, M
Holder, DS
机构
[1] UCL, Middlesex Hosp, Dept Clin Neurophysiol, London W1T 3AA, England
[2] Middlesex Univ, Sch Hlth Environm & Biol Sci, London N19 5ND, England
[3] UCL Hosp NHS Trust, Dept Med Phys & Bioengn, London WC1E 6JA, England
关键词
EIT; hardware; design; performance;
D O I
10.1088/0967-3334/23/1/314
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The UCLH Mark 1b is a portable EIT system that can address up to 64 electrodes, which has been designed for imaging brain function with scalp electrodes. It employs a single impedance-measuring circuit and multiplexer so that electrode combinations may be addressed flexibly using software. It operates in the relatively low frequency band between 225 Hz and 77 kHz, as lower frequencies produce larger changes during brain activity, and has a videocassette-sized headbox on a lead 10 m long, connected to a base box the size of a video recorder, and notebook PC, so that recordings may be made in ambulant subjects. Its performance was assessed using a resistor-capacitor network, and two saline-filled tanks-a cylindrical Perspex one and a latex one which contained a human skull. System signal-to-noise ratio was better than 50 dB and the maximum reciprocity error less than 10% for most frequencies. The CMMR was better than 80 dB at 38 kHz and a sponge, 20 mm across, which caused a local 12% impedance increase, was correctly localized in images. This suggests that the system has adequate performance to image impedance changes of 5-50% known to occur in the brain during normal activity, epilepsy or stroke; clinical trials to image these conditions are in progress.
引用
收藏
页码:149 / 158
页数:10
相关论文
共 15 条
[1]   Improvement of the positional accuracy of EIT images of the head using a Lagrange multiplier reconstruction algorithm with diametric excitation [J].
Bayford, RH ;
Boone, KG ;
Hanquan, Y ;
Holder, DS .
PHYSIOLOGICAL MEASUREMENT, 1996, 17 :A49-A57
[2]   Current approaches to analogue instrumentation design in electrical impedance tomography [J].
Boone, KG ;
Holder, DS .
PHYSIOLOGICAL MEASUREMENT, 1996, 17 (04) :229-247
[3]  
BROWN BH, 2001, BIOMEDICAL APPL EIT
[4]  
BROWN BH, 1999, MED PHYSICS BIOMEDIC
[5]  
Cusick G., 1994, INNOV TECH BIOL MED, V15, P40
[6]   A COLE PHANTOM FOR EIT [J].
GRIFFITHS, H .
PHYSIOLOGICAL MEASUREMENT, 1995, 16 :A29-A38
[7]   Assessment and calibration of a low-frequency system for electrical impedance tomography (EIT), optimized for use in imaging brain function in ambulant human subjects [J].
Holder, DS ;
González-Correa, CA ;
Tidswell, T ;
Gibson, A ;
Cusick, G ;
Bayford, RH .
ELECTRICAL BIOIMPEDANCE METHODS: APPLICATIONS TO MEDICINE AND BIOTECHNOLOGY, 1999, 873 :512-519
[8]  
HOLDER DS, 1994, INNOV TECH BIOL MED, V15, P33
[9]   EIT data noise evaluation in the clinical environment [J].
Meeson, S ;
Blott, BH ;
Killingback, ALT .
PHYSIOLOGICAL MEASUREMENT, 1996, 17 :A33-A38
[10]  
NEWELL JC, 2001, BIOMEDICAL APPL EIT