Rotary blood pump flow spontaneously increases during exercise under constant pump speed: Results of a chronic study

被引:58
作者
Akimoto, T
Yamazaki, K
Litwak, P
Litwak, KN
Tagusari, O
Mori, T
Antaki, JF
Kameneva, MV
Watach, MJ
Umezu, M
Tomioka, J
Kormos, RL
Koyanagi, H
Griffith, BP
机构
[1] Univ Pittsburgh, McGowan Ctr, Pittsburgh, PA 15219 USA
[2] Tokyo Womens Med Coll, Tokyo 162, Japan
[3] Sun Med Technol Res Corp, Nagano, Japan
[4] Waseda Univ, Tokyo, Japan
关键词
implantable centrifugal blood pump; chronic experiment; exercise; hemodynamics; pump control; heart rate;
D O I
10.1046/j.1525-1594.1999.06426.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Many types of rotary blood pumps and pump control methods have recently been developed with the goal of clinical use. From experiments, we know that pump flow spontaneously increases during exercise without changing pump control parameters. The purpose of this study was to determine the hemodynamics associated with the long-term observation of calves implanted with centrifugal blood pumps (EVAHEART, Sun Medical Technology Research Corporation, Nagano, Japan). Two healthy female Jersey calves were implanted with devices in the left thoracic cavity. A total of 22 treadmill exercise tests were performed after the 50th postoperative day. During exercise, the following parameters were compared with conditions at rest: heart rate, blood pressure, central venous oxygen saturation (SvO(2)), pump speed, and pump flow. The pump flow in a cardiac cycle was analyzed by separating the systole and diastole. Compared to the base data, statistically significant differences were found in the following interrelated parameters: the heart rate (66.8 +/- 5.2 vs. 196 +/- 9.7 bpm), mean pump flow (4.8 +/- 0.2 vs. 7.0 +/- 0.3 L/min), and volume of pump now in diastole (26.9 +/- 1.8 vs. 13.5 +/- 2.5 ml). During exercise, the volume of pump flow in systole was 3 times larger than that measured in diastole. Blood pressure, SvO(2), and pump speed did not change significantly from rest to exercise. These results suggested that the mean pump flow depends on the systolic pump flow. Therefore, the increase in the mean pump flow during exercise under constant pump speed was caused by an increase in the heart rate.
引用
收藏
页码:797 / 801
页数:5
相关论文
共 16 条
[1]   Development of the Nimbus/Pittsburgh axial flow left ventricular assist system [J].
Butler, K ;
Thomas, D ;
Antaki, J ;
Borovetz, H ;
Griffith, B ;
Kameneva, M ;
Kormos, R ;
Litwak, P .
ARTIFICIAL ORGANS, 1997, 21 (07) :602-610
[2]  
Delgado R, 1998, ASAIO J, V44, P299
[3]  
Foray A, 1996, CIRCULATION, V94, P222
[4]  
GOLDING LR, 1980, T AM SOC ART INT ORG, V26, P251
[5]  
Guyton AC, 1996, TXB MED PHYSL, P253
[6]   SYSTEM CONSIDERATIONS FAVORING ROTARY ARTIFICIAL HEARTS WITH BLOOD-IMMERSED BEARINGS [J].
JARVIK, RK .
ARTIFICIAL ORGANS, 1995, 19 (07) :565-570
[7]   Effects of exercise during long-term support with a left ventricular assist device - Results of the experience with left ventricular assist device with exercise (EVADE) pilot trial [J].
Jaski, BE ;
Kim, J ;
Maly, RS ;
Branch, KR ;
Adamson, R ;
Favrot, LK ;
Smith, SC ;
Dembitsky, WP .
CIRCULATION, 1997, 95 (10) :2401-2406
[8]   Controller for an axial flow blood pump [J].
Konishi, H ;
Antaki, JF ;
Amin, DV ;
Boston, JR ;
Kerrigan, JP ;
Mandarino, WA ;
Litwak, P ;
Yamazaki, K ;
Macha, M ;
Butler, KC ;
Borovetz, HS ;
Kormos, RL .
ARTIFICIAL ORGANS, 1996, 20 (06) :618-620
[9]  
KONISHI H, 1994, T AM SOC ART INT ORG, V40, pM299
[10]  
NISHIMURA K, 1997, T AM SOC ART INT ORG, V43, pM553