On-line statistical processing of radiation detector pulse trains with time-varying count rates

被引:10
作者
Apostolopoulos, G. [1 ]
机构
[1] Natl Ctr Sci Res Demokritos, Inst Nucl Technol & Radiat Protect, GR-15310 Aghia Parakevi Attikis, Greece
关键词
Statistical signal processing; Radiation detectors; Random pulse trains; Radiation monitoring; Rate meter algorithms; Poisson regression; Maximum-likelihood estimation; Generalized likelihood ratio test;
D O I
10.1016/j.nima.2008.06.050
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Statistical analysis is of primary importance for the correct interpretation of nuclear measurements, due to the inherent random nature of radioactive decay processes. This paper discusses the application of statistical signal processing techniques to the random pulse trains generated by radiation detectors. The aims of the presented algorithms are: (i) continuous, on-line estimation of the underlying time-varying count rate theta(t) and its first-order derivative d theta/dt; (ii) detection of abrupt changes in both of these quantities and estimation of their new value after the change point. Maximum-likelihood techniques, based on the Poisson probability distribution, are employed for the on-line estimation of theta and d theta/dt. Detection of abrupt changes is achieved on the basis of the generalized likelihood ratio statistical test. The properties of the proposed algorithms are evaluated by extensive simulations and possible applications for on-line radiation monitoring are discussed. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:464 / 473
页数:10
相关论文
共 53 条