The hemodynamic effects of double-orifice valve repair for mitral regurgitation: a 3D computational model

被引:59
作者
Maisano, F
Redaelli, A
Pennati, G
Fumero, R
Torracca, L
Alfieri, O
机构
[1] Hosp San Raffaele, IRCCS, Dept Cardiac Surg, I-20132 Milan, Italy
[2] Politecn Milan, CEBITEC, Milan, Italy
关键词
valve repair; computer modelling; mitral regurgitation; echo-Doppler;
D O I
10.1016/S1010-7940(99)00071-8
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: A 3D computational model has been implemented for the evaluation of the hemodynamics of the double orifice repair. Critical issues for surgical decision making and echo-Doppler evaluation of the results of the procedure are investigated. Methods: A parametric 3D computational model of the double-orifice mitral valve based on the finite elements model has been constructed from clinical data. Nine different geometries were investigated, corresponding to three total inflow areas (1.5, 2.25 and 3 cm(2)) and to three orifice configurations (two equal orifices, two orifices of different areas, i.e. one twice as much the other one, and a single orifice). The simulations were performed in transit; the fluid was initially quiescent and was accelerated to the maximum flow rate with a cubic function. For each case, some characteristic values of velocity and pressure were determined: velocities were calculated downstream of each orifice, at the centre of it (V(cen1), V(cen2)). The maximum velocity was also determined for each orifice (V(max1), V(max2)). Maximum pressure drops (Delta p(max)) across the valve were compared with the estimations (Delta p(Bernoulli)) based on the Bernoulli formula (4 V(2)). Results; In each simulation, no notable difference was observed between V(cen1) and V(cen2), and between V(max1) and V(max2), regardless of the valve configuration. Maximum velocity and Delta p(max) were related to the total orifice area and were not influenced by the orifice configuration. Delta p(Bernoulli) calculated with V(max) was well correlated with the Delta p(max) obtained throughout the simulations (y = 0.9126x + 0.3464, r = 0.996); on the contrary the pressure drops estimated using V(cen) underestimated (y = 0.6757x + 0.3073, r = 0.999) the actual pressure drops. Conclusions: The hemodynamic behaviour of a double orifice mitral valve does not differ from that of a physiological valve of same total area: pressure drops and flow velocity across the valve are not influenced by the configuration of the valve. Echo Doppler estimation of the maximum velocities is a reliable method for the calculation of pressure gradients across the repaired valve. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:419 / 425
页数:7
相关论文
共 11 条
[1]   DOUBLE-ORIFICE MITRAL-VALVE - A STUDY OF 27 POSTMORTEM CASES WITH DEVELOPMENTAL, DIAGNOSTIC AND SURGICAL CONSIDERATIONS [J].
BANORODRIGO, A ;
VANPRAAGH, S ;
TROWITZSCH, E ;
VANPRAAGH, R .
AMERICAN JOURNAL OF CARDIOLOGY, 1988, 61 (01) :152-160
[2]   Use of computational fluid dynamics in the design of surgical procedures: Application to the study of competitive flows in cavopulmonary connections [J].
deLeval, MR ;
Dubini, G ;
Migliavacca, F ;
Jalali, H ;
Camporini, G ;
Redington, A ;
Pietrabissa, R .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1996, 111 (03) :502-510
[3]  
*FLUID DYN INT, 1993, FIDAP THER MAN
[4]  
Ge S, 1996, J Am Soc Echocardiogr, V9, P700, DOI 10.1016/S0894-7317(96)90067-X
[5]  
HOLEN J, 1976, ACTA MED SCAND, V199, P455
[6]   The edge-to-edge technique: a simplified method to correct mitral insufficiency [J].
Maisano, F ;
Torracca, L ;
Oppizzi, M ;
Stefano, PL ;
D'Addario, G ;
La Canna, G ;
Zogno, M ;
Alfieri, O .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1998, 13 (03) :240-245
[7]   Valve repair for traumatic tricuspid regurgitation [J].
Maisano, F ;
Lorusso, R ;
Sandrelli, L ;
Torracca, L ;
Coletti, G ;
LaCanna, G ;
Alfieri, O .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1996, 10 (10) :867-873
[8]   Partial left ventriculectomy and mitral valve repair for end-stage congestive heart failure [J].
McCarthy, JF ;
McCarthy, PM ;
Starling, RC ;
Smedira, NG ;
Scalia, GM ;
Wong, J ;
Kasirajan, V ;
Goormastic, M ;
Young, JB .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1998, 13 (04) :337-343
[9]   Computational analysis of the ductus venosus fluid dynamics based on Doppler measurements [J].
Pennati, G ;
Redaelli, A ;
Bellotti, M ;
Ferrazzi, E .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1996, 22 (08) :1017-1029
[10]   Computational evaluation of intraventricular pressure gradients based on a fluid-structure approach [J].
Redaelli, A ;
Montevecchi, FM .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1996, 118 (04) :529-537