Safety and effectiveness of placement of pacemaker and defibrillator leads in the axillary vein guided by contrast venography

被引:60
作者
Ramza, BM
Rosenthal, L
Hui, R
Nsah, E
Savader, S
Lawrence, JH
Tomaselli, G
Berger, R
Brinker, J
Calkins, H
机构
[1] JOHNS HOPKINS UNIV,SCH MED,DEPT MED,BALTIMORE,MD 21287
[2] JOHNS HOPKINS UNIV,SCH MED,DEPT RADIOL,BALTIMORE,MD 21287
关键词
D O I
10.1016/S0002-9149(97)00542-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Despite evidence of an increased incidence of lead fracture, the infraclavicular subclavian approach remains the dominant approach for placement of pacemaker and implantable defibrillator leads. Although this complication can be prevented by lead placement in the cephalic vein or by recently described approaches for lead placement in the axillary vein, these approaches have not gained widespread acceptance. The purpose of this study was to evaluate the safety and efficacy of an alternative technique for lead placement that uses contrast-guided venipuncture of the axillary vein with a 5Fr micropuncture introducer set, A fetal of 50 patients underwent an attempt at placement of pacemaker or implantable defibrillator leads via the axillary vein using this new technique. Patients were randomized into 2 groups based on whether the initial attempt at axillary vein access was performed medial or lateral to the rib cage margin. Leed placement was successfully accomplished in 49 of the 50 patients using this technique. initial success was achieved in each of 25 patients randomized to the medial approach compared with 18 of 24 patients randomized to the lateral approach to the axillary vein (75%). In each of the 6 patients in whom the initial technique failed, lead placement was subsequently achieved with the medial approach. In addition to a higher initial success rate, the medial approach was determined to be preferable as evidenced by a shorter lead placement time, a smaller number of contrast injections, and a reduced requirement for additional micropuncture guidewires. There were no major complications associated with either approach. Contrast-guided venipuncture of the axillary vein is a safe and effective approach to placement of endocardial leads. (C) 1997 by Excerpta Medica, Inc.
引用
收藏
页码:892 / 896
页数:5
相关论文
共 30 条
  • [1] AGGARWAL RK, 1995, BRIT HEART J, V73, P571
  • [2] THE PINCH-OFF SIGN - A WARNING OF IMPENDING PROBLEMS WITH PERMANENT SUBCLAVIAN CATHETERS
    AITKEN, DR
    MINTON, JP
    [J]. AMERICAN JOURNAL OF SURGERY, 1984, 148 (05) : 633 - 636
  • [3] LEAD FRACTURE IN PACEMAKER PATIENTS
    ALT, E
    VOLKER, R
    BLOMER, H
    [J]. THORACIC AND CARDIOVASCULAR SURGEON, 1987, 35 (02) : 101 - 104
  • [4] Multicenter experience with a pectoral unipolar implantable cardioverter-defibrillator
    Bardy, GH
    Yee, R
    Jung, W
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1996, 28 (02) : 400 - 410
  • [5] SURVEY OF CARDIAC PACING IN THE UNITED-STATES IN 1989
    BERNSTEIN, AD
    PARSONNET, V
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1992, 69 (04) : 331 - 338
  • [6] Bonavita G, 1990, PACE, V13, P551
  • [7] ENDOCARDIAL PACING LEADS - THE GOOD, THE BAD, AND THE UGLY
    BRINKER, JA
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (05): : 953 - 954
  • [8] SAFE INTRODUCER TECHNIQUE FOR PACEMAKER LEAD IMPLANTATION
    BYRD, CL
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (03): : 262 - 267
  • [9] CLINICAL-EXPERIENCE WITH THE EXTRATHORACIC INTRODUCER INSERTION TECHNIQUE
    BYRD, CL
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (09): : 1781 - 1784
  • [10] EARLY COMPLICATIONS AFTER DUAL-CHAMBER VERSUS SINGLE CHAMBER PACEMAKER IMPLANTATION
    CHAUHAN, A
    GRACE, AA
    NEWELL, SA
    STONE, DL
    SHAPIRO, LM
    SCHOFIELD, PM
    PETCH, MC
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (11): : 2012 - 2015