Experience with an ultrasound stethoscope

被引:51
作者
Vourvouri, EC [1 ]
Poldermans, D [1 ]
De Sutter, J [1 ]
Sozzi, FB [1 ]
Izzo, P [1 ]
Roelandt, JRTC [1 ]
机构
[1] Erasmus Med Ctr, Dept Cardiol, Thoraxctr, NL-3015 GD Rotterdam, Netherlands
关键词
D O I
10.1067/mje.2002.115268
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. To test the diagnostic potential of the SonoHeart, a battery-powered hand-held ultrasound imaging device, in an outpatient clinic setting. Methods: A total of 114 patients with a variety of cardiac diseases were examined by 2 independent cardiologists with the hand-held device using the standard echocardiographic system (SE) as a reference. Global right ventricular (RV) and left ventricular (LV) function (scored as normal, mildly to moderately, or severely reduced) and internal cavity dimensions were assessed. Regional wall motion of 6 segments using a 2-point score (1 = normal wall motion, 2 = abnormal wall motion) was evaluated in 34 patients on-line. Results. There was a good agreement between the 2 imaging devices for evaluation of global LV (93%) and RV function (99%), regional wall motion (90%), dimensions of the LV (99%) and the RV (99%), and the left (96%) and right atria (99%). Furthermore, SonoHeart identified hypertrophic cardiomyopathy, pericardial effusion, and abnormalities of valves. Conclusion. The SonoHeart device allows rapid and accurate diagnosis, whenever needed in the outpatient clinic.
引用
收藏
页码:80 / 85
页数:6
相关论文
共 11 条
  • [1] THE LIMITED ECHOCARDIOGRAM - A MODIFICATION OF STANDARD ECHOCARDIOGRAPHY FOR USE IN THE ROUTINE EVALUATION OF PATIENTS WITH SYSTEMIC HYPERTENSION
    BLACK, HR
    WELTIN, G
    JAFFE, CC
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1991, 67 (11) : 1027 - 1030
  • [2] Personal ultrasound imager: Abdominal aortic aneurysm screening
    Bruce, CJ
    Spittell, PC
    Montgomery, SC
    Bailey, KR
    Tajik, AJ
    Seward, JB
    [J]. JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2000, 13 (07) : 674 - 679
  • [3] Fleiss JL, 1981, STAT METHODS RATES P
  • [4] Indications for limited echocardiographic imaging: A mathematical model
    Kimura, BJ
    DeMaria, AN
    [J]. JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2000, 13 (09) : 855 - 861
  • [5] Accuracy and cost-effectiveness of single-view echocardiographic screening for suspected mitral valve prolapse
    Kimura, BJ
    Scott, R
    Willis, CL
    DeMaria, AN
    [J]. AMERICAN JOURNAL OF MEDICINE, 2000, 108 (04) : 331 - 333
  • [6] LIGTVOET C, 1978, Ultrasound in Medicine and Biology, V4, P91, DOI 10.1016/0301-5629(78)90033-9
  • [7] ROELANDT J, 1978, Ultrasound in Medicine and Biology, V4, P93, DOI 10.1016/0301-5629(78)90034-0
  • [8] THE ULTRASOUND CARDIOSCOPE - A HAND-HELD SCANNER FOR REAL-TIME CARDIAC IMAGING
    ROELANDT, J
    BOM, K
    HUGENHOLTZ, PG
    [J]. JOURNAL OF CLINICAL ULTRASOUND, 1980, 8 (03) : 221 - 225
  • [9] Limited echocardiography for hypertensive left ventricular hypertrophy
    Sheps, SG
    Frohlich, ED
    [J]. HYPERTENSION, 1997, 29 (02) : 560 - 563
  • [10] Shub C, 1995, J Am Soc Echocardiogr, V8, P280, DOI 10.1016/S0894-7317(05)80038-0