摘要
Assay conditions of a plasma-based ecarin clotting time (ECT) were evaluated and the precision of the ECT in monitoring plasma levels of r-hirudin assessed. The snake venom enzyme ecarin converts prothrombin to meizothrombin possessing only weak coagulant but strong esterase activity. In r-hirudin containing plasma samples, meizo thrombin is rapidly neutralized by r-hirudin resulting in a dose-dependent prolongation of the clotting times. Among the different assay conditions tested, addition of 50 mu l of ecarin (4 U/ml) to 100 mu l of undiluted citrate-anticoagulated plasma gave optimal results with respect to precision and reproducibility. The measuring range of the ECT performed in this way is about 0.02 - 5.0 mu g/ml r-hirudin. In vitro studies performed on r-hirudin-spiked plasma samples of 50 healthy individuals demonstrate remarkably low interindividual differences in r-hirudin responsiveness as indicated by CV-values below 5% and 7% at r-hirudin concentrations between 0 - 3 mu g/ml and 4 - 5 mu g/ml, respectively. The specificity for r-hirudin of the ECT is further demonstrated by the strong correlation (r=0.94) between the results of a chromogenic assay and the ECT-measured r-hirudin concentrations obtained on 67 ex vivo blood samples. Depending on the concentration of r-hirudin the ECT is sensitive to plasma levels of prothrombin. In the absence of r-hirudin the critical prothrombin concentration was found to be 20% but increasing to 60% in the presence of 2.0 mu g/ml r-hirudin. A fibrinogen concentration of 50 mg/dl was found to be the minimal concentration independent of the r-hirudin concentration. The precision of the ECT in measuring plasma levels of r-hirudin is not influenced by treatment with heparin, aprotinin or oral anticoagulants. The data demonstrate that the ECT is a rapid and easily performable clotting assay which allows accurate measurement of r-hirudin plasma levels. ECT-monitored hirudin treatment will help to establish optimal dose regimens that are more efficacious but still as safe as heparin. (C) 1997 Elsevier Science Ltd.
引用
收藏
页码:373 / 383
页数:11
相关论文
共 15 条
  • [1] HIRUDIN IN ACUTE MYOCARDIAL-INFARCTION - SAFETY REPORT FROM THE THROMBOLYSIS AND THROMBIN INHIBITION IN MYOCARDIAL-INFARCTION (TIMI)-9A TRIAL
    ANTMAN, EM
    [J]. CIRCULATION, 1994, 90 (04) : 1624 - 1630
  • [2] GREINACHER A, 1996, ANN HEMATOL, V72, pA92
  • [3] EVALUATION OF AN AUTOMATED CHROMOGENIC SUBSTRATE ASSAY FOR THE RAPID-DETERMINATION OF HIRUDIN IN PLASMA
    HAFNER, G
    FICKENSCHER, K
    FRIESEN, HJ
    RUPPRECHT, HJ
    KONHEISER, U
    EHRENTHAL, W
    LOTZ, J
    PRELLWITZ, W
    [J]. THROMBOSIS RESEARCH, 1995, 77 (02) : 165 - 173
  • [4] SAFETY OBSERVATIONS FROM THE PILOT PHASE OF THE RANDOMIZED R-HIRUDIN FOR IMPROVEMENT OF THROMBOLYSIS (HIT-III) STUDY - A STUDY OF THE ARBEITSGEMEINSCHAFT-LEITENDER-KARDIOLOGISCHER-KRANKENHAUSARZTE (ALKK)
    NEUHAUS, KL
    VONESSEN, R
    TEBBE, U
    JESSEL, A
    HEINRICHS, H
    MAURER, W
    DORING, W
    HARMJANZ, D
    KOTTER, V
    KALHAMMER, E
    SIMON, H
    HORACEK, T
    [J]. CIRCULATION, 1994, 90 (04) : 1638 - 1642
  • [5] MECHANISMS OF ALPHA-THROMBIN AND BETA-THROMBIN-E FORMATION - USE OF ECARIN FOR ISOLATION OF MEIZOTHROMBIN-1
    NOVOA, E
    SEEGERS, WH
    [J]. THROMBOSIS RESEARCH, 1980, 18 (05) : 657 - 668
  • [6] NOWAK G, 1993, THROMB HAEMOSTASIS, V69, P1306
  • [7] Quantitative determination of hirudin in blood and body fluids
    Nowak, G
    Bucha, E
    [J]. SEMINARS IN THROMBOSIS AND HEMOSTASIS, 1996, 22 (02) : 197 - 202
  • [8] NURMOHAMED MT, 1992, THROMB HAEMOSTASIS, V72, P685
  • [9] SOLANO C, 1990, THROMB HAEMOSTASIS, V64, P353
  • [10] VANDENBOS AA, 1993, CIRCULATION, V88, P58