Array biosensor for simultaneous identification of bacterial, viral, and protein analytes

被引:243
作者
Rowe, CA
Tender, LM
Feldstein, MJ
Golden, JP
Scruggs, SB
MacCraith, BD
Cras, JJ
Ligler, FS
机构
[1] USN, Res Lab, Ctr Biomol Sci & Engn, Washington, DC 20375 USA
[2] Geocenters Inc, Rockville, MD 20852 USA
[3] Dublin City Univ, Dept Phys, Dublin 9, Ireland
关键词
D O I
10.1021/ac981425v
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The array biosensor was fabricated to analyze multiple samples simultaneously for multiple analytes. The sensor utilized a standard sandwich immunoassay format: Antigen-specific "capture" antibodies were immobilized in a patterned array on the surface of a planar waveguide and bound analyte was subsequently detected using fluorescent tracer antibodies. This study describes the analysis of 126 blind samples for the presence of three distinct classes of analytes. To address potential complications arising from using a mixture of tracer antibodies in the multianalyte assay, three single-analyte assays were run in parallel with a multianalyte assay. Mixtures of analytes were also assayed to demonstrate the sensor's ability to detect more than a single species at a time. The array sensor was capable of detecting viral, bacterial, and protein analytes using a facile 14-min assay with sensitivity levels approaching those of standard ELISA methods. Limits of detection for Bacillus globigii, MS2 bacteriophage, and staphylococcal enterotoxin B (SEB) were 10(5) cfu/mL, 10(7) pfu/mL, and 10 ng/mL, respectively, The array biosensor also analyzed multiple samples simultaneously and detected mixtures of the different types of analytes in the multianalyte format.
引用
收藏
页码:3846 / 3852
页数:7
相关论文
共 25 条
[1]   Fiber-optic evanescent wave biosensor for the detection of oligonucleotides [J].
Abel, AP ;
Weller, MG ;
Duveneck, GL ;
Ehrat, M ;
Widmer, HM .
ANALYTICAL CHEMISTRY, 1996, 68 (17) :2905-2912
[2]   A fiber optic biosensor for multianalyte detection: importance of preventing fluorophore aggregation [J].
Bakaltcheva, IB ;
Shriver-Lake, LC ;
Ligler, FS .
SENSORS AND ACTUATORS B-CHEMICAL, 1998, 51 (1-3) :46-51
[3]   Surface plasmon resonance multisensing [J].
Berger, CEH ;
Beumer, TAM ;
Kooyman, RPH ;
Greve, J .
ANALYTICAL CHEMISTRY, 1998, 70 (04) :703-706
[4]   USE OF THIOL-TERMINAL SILANES AND HETEROBIFUNCTIONAL CROSSLINKERS FOR IMMOBILIZATION OF ANTIBODIES ON SILICA SURFACES [J].
BHATIA, SK ;
SHRIVERLAKE, LC ;
PRIOR, KJ ;
GEORGER, JH ;
CALVERT, JM ;
BREDEHORST, R ;
LIGLER, FS .
ANALYTICAL BIOCHEMISTRY, 1989, 178 (02) :408-413
[5]   Step-and-repeat photopatterning of protein features using caged/biotin-BSA: Characterization and resolution [J].
Blawas, AS ;
Oliver, TF ;
Pirrung, MC ;
Reichert, WM .
LANGMUIR, 1998, 14 (15) :4243-4250
[6]   Optical immunoprobe development for multiresidue monitoring in water [J].
Brecht, A ;
Klotz, A ;
Barzen, C ;
Gauglitz, G ;
Harris, RD ;
Quigley, GR ;
Wilkinson, JS ;
Sztajnbok, P ;
Abuknesha, R ;
Gascón, J ;
Oubiña, A ;
Barceló, D .
ANALYTICA CHIMICA ACTA, 1998, 362 (01) :69-79
[7]  
EIGLER FS, 1991, Patent No. 5077210
[8]   FLUORESCENCE SPECTROSCOPY AND ITS APPLICATION TO A NEW GENERATION OF HIGH-SENSITIVITY, MULTI-MICROSPOT, MULTIANALYTE, IMMUNOASSAY [J].
EKINS, R ;
CHU, F ;
BIGGART, E .
CLINICA CHIMICA ACTA, 1990, 194 (01) :91-114
[9]  
EKINS R, 1993, CLIN CHEM, V39, P369
[10]   Array Biosensor: Optical and Fluidics Systems [J].
Feldstein, Mark J. ;
Golden, Joel P. ;
Rowe, Chris A. ;
MacCraith, Brian D. ;
Ligler, Frances S. .
BIOMEDICAL MICRODEVICES, 1999, 1 (02) :139-153