Determination of Creatinine in Human Urine with Flow Injection Tandem Mass Spectrometry

被引:19
作者
Niesser, Mareile [1 ]
Koletzko, Berthold [1 ]
Peissner, Wolfgang [1 ]
机构
[1] Univ Munich, Med Ctr, Dr von Hauner Childrens Hosp, Div Metab & Nutr Med, DE-80337 Munich, Germany
基金
欧盟第七框架计划;
关键词
Creatinine; Human urine; Tandem mass spectrometry; Flow injection analysis; Qualifier ion monitoring; QUANTITATIVE-ANALYSIS; ENZYMATIC METHOD; QUANTIFICATION; METABOLISM; PLASMA; SERUM; ELECTROPHORESIS; MS/MS; ACID; HPLC;
D O I
10.1159/000342774
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background/Aims: Excretion of urinary compounds in spot urine is often estimated relative to creatinine. For the growing number of liquid chromatography-tandem mass spectrometry (LC-MS/MS) assays of urine-excreted molecules, a fast and accurate method for determination of creatinine is needed. Methods: A high-throughput flow injection tandem mass spectrometry method for exact quantitation of creatinine in urine has been developed and validated. Sample preparation used only two-step dilution for protein precipitation and matrix dilution. Flow injection analysis without chromatographic separation allowed for total run times of 1 min per sample. Creatinine concentrations were quantitated using stable isotope dilution tandem mass spectrometry. Selectivity and coelution-free quantitation were assured by qualifier ion monitoring. Results: Method validation revealed excellent injection repeatability of 1.0% coefficient of variation (CV), intraday precision of 1.2% CV and interday precision of 2.4% CV. Accuracy determined from standard addition experiments was 106.1 +/- 3.8%. The linear calibration range was adapted to physiological creatinine concentrations. Comparison of quantitation results with a routinely used method (Jaffe colorimetric assay) proved high agreement (R-2 = 0.9102). Conclusions: The new method is a valuable addition to the toolbox of LC-MS/MS laboratories where excretion of urinary compounds is studied. The 'dilute and shoot' approach to isotope dilution tandem mass spectrometry makes the new method highly accurate as well as cost-and time-efficient. Copyright (C) 2012 S. Karger AG, Basel
引用
收藏
页码:314 / 321
页数:8
相关论文
共 40 条
[1]  
[Anonymous], 2001, GUID IND BIOAN METH
[2]  
[Anonymous], 1886, physiol. Chem., DOI DOI 10.1515/BCHM1.1886.10.5.391
[3]   Urine-creatinine concentration as a marker of urine dilution: Reflections using a cohort of 45,000 samples [J].
Arndt, T. .
FORENSIC SCIENCE INTERNATIONAL, 2009, 186 (1-3) :48-51
[4]   Urinary creatinine concentrations in the US population: Implications for urinary biologic monitoring measurements [J].
Barr, DB ;
Wilder, LC ;
Caudill, SP ;
Gonzalez, AJ ;
Needham, LL ;
Pirkle, JL .
ENVIRONMENTAL HEALTH PERSPECTIVES, 2005, 113 (02) :192-200
[5]   A new liquid chromatography/mass spectrometry method for 4-(methylnitrosamino)-1-(3-pyridyl)-1-butanol (NNAL) in urine [J].
Bhat, Showket H. ;
Gelhaus, Stacy L. ;
Mesaros, Clementina ;
Vachani, Anil ;
Blair, Ian A. .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2011, 25 (01) :115-121
[6]   Analysis of creatine and creatinine in urine by capillary electrophoresis [J].
Burke, DG ;
MacLean, PG ;
Walker, RA ;
Dewar, PJ ;
Smith-Palmer, T .
JOURNAL OF CHROMATOGRAPHY B, 1999, 732 (02) :479-485
[7]   A biochemical perspective on the use of tandem mass spectrometry for newborn screening and clinical testing [J].
Chace, DH ;
Kalas, TA .
CLINICAL BIOCHEMISTRY, 2005, 38 (04) :296-309
[8]  
CHACE DH, 1993, CLIN CHEM, V39, P66
[9]   Limits of suspicion, recognition and confirmation as concepts that account for the confirmation transitions at the detection limit for quantification by liquid chromatography-tandem mass spectrometry [J].
Delatour, Thierry ;
Mottier, Pascal ;
Gremaud, Eric .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1169 (1-2) :103-110
[10]  
GUDER WG, 1986, J CLIN CHEM CLIN BIO, V24, P889