Effect of respiratory gating on reducing lung motion artifacts in PET imaging of lung cancer

被引:236
作者
Nehmeh, SA
Erdi, YE
Ling, CC
Rosenzweig, KE
Squire, OD
Braban, LE
Ford, E
Sidhu, K
Mageras, GS
Larson, SM
Humm, JL
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10021 USA
[2] Mem Sloan Kettering Canc Ctr, Dept Radiat Oncol, New York, NY 10021 USA
[3] Mem Sloan Kettering Canc Ctr, Div Nursing, New York, NY 10021 USA
[4] Mem Sloan Kettering Canc Ctr, Nucl Med Serv, Dept Radiol, New York, NY 10021 USA
关键词
positron emission tomography; gated-PET; respiration; standard uptake value (SUV); planning target volume (PTV); F-18-fluorodeoxy-glucose (FDG);
D O I
10.1118/1.1448824
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Positron emission tomography (PET) has shown an increase in both sensitivity and specificity over computed tomography (CT) in lung cancer. However, motion artifacts in the F-18 fluorodioxydoglucose (FDG) PET images caused by respiration persists to be an important factor in degrading PET image quality and quantitication. Motion artifacts lead to two major effects: First, it affects the accuracy of quantitation, producing a reduction of the measured standard uptake value (SUV). Second, the apparent lesion volume is overestimated. Both impact upon the usage of PET images for radiation treatment planning. The first affects the visibility, or contrast, of the lesion. The second results in an increase in the planning target volume, and consequently a greater radiation dose to the normal tissues. One way to compensate for this effect is by applying a multiple-frame capture technique. The PET data are then acquired in synchronization with the respiratory motion. Reduction in smearing due to gating was investigated in both phantoms and patient studies. Phantom studies showed a dependence of the reduction in smearing on the lesion size, the motion amplitude, and the number of bins used for data acquisition. These studies also showed an improvement in the target-to-background ratio, and a more accurate measurement of the SUV. When applied to one patient, respiratory gating showed a 28% reduction in the total lesion volume. and a 56.5% increase in the SUV. This study was conducted as a proof of principle that a gating technique can effectively reduce motion artifacts in PET image acquisition. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:366 / 371
页数:6
相关论文
共 16 条
  • [1] Brink I, 2001, J NUCL MED, V42, P591
  • [2] DEGRADO TR, 1994, J NUCL MED, V35, P1398
  • [3] Use of PET to monitor the response of lung cancer to radiation treatment
    Erdi, YE
    Macapinlac, H
    Rosenzweig, KE
    Humm, JL
    Larson, SM
    Erdi, AK
    Yorke, ED
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 2000, 27 (07) : 861 - 866
  • [4] Erdi YE, 1997, CANCER, V80, P2505, DOI 10.1002/(SICI)1097-0142(19971215)80:12+<2505::AID-CNCR24>3.0.CO
  • [5] 2-F
  • [6] INOUE T, 1995, J NUCL MED, V36, P788
  • [7] KEYES JW, 1995, J NUCL MED, V36, P1836
  • [8] Respiration gated radiotherapy treatment: A technical study
    Kubo, HD
    Hill, BC
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (01) : 83 - 91
  • [9] Breathing-synchronized radiotherapy program at the University of California Davis Cancer Center
    Kubo, HD
    Len, PM
    Minohara, S
    Mostafavi, H
    [J]. MEDICAL PHYSICS, 2000, 27 (02) : 346 - 353
  • [10] Larson Steven M., 1999, Clin Positron Imaging, V2, P159, DOI 10.1016/S1095-0397(99)00016-3