Thermography and colour duplex ultrasound assessments of arterio-venous fistula function in renal patients

被引:9
作者
Allen, J [1 ]
Oates, CP
Chishti, AD
Ahmed, HAM
Talbot, D
Murray, A
机构
[1] Freeman Rd Hosp, Dept Reg Med Phys, Newcastle Upon Tyne NE7 7DN, Tyne & Wear, England
[2] Freeman Rd Hosp, Dept Anesthesia, Newcastle Upon Tyne NE7 7DN, Tyne & Wear, England
[3] Freeman Rd Hosp, Renal Med & Transplant Ctr, Newcastle Upon Tyne NE7 7DN, Tyne & Wear, England
关键词
arterio-venous fistula; infrared thermography; ultrasound; blood flow; vascular steal; end stage renal disease; haemodialysis;
D O I
10.1088/0967-3334/27/1/005
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Vascular and clinical assessments of arterio-venous fistula (AVF) function and access are important in patients undergoing or preparing to undergo renal dialysis. Objective assessment techniques include colour duplex ultrasound and more recently medical infrared thermography. Ideally, these should help assess problems relating to fistula failure or to vascular steal from the hand which can result from excessive fistula blood flow. The clinical value of thermography, as yet, has not been assessed for this patient group. The aims of this study were therefore to investigate the relationships between thermography skin temperature measurement and (a) quantitative ultrasound measurement of AVF blood flow, and (b) qualitative clinical assessment of vascular steal from the hands. Fifteen adult patients underwent thermal imaging of the upper limbs, colour duplex ultrasound to derive AVF blood flow from brachial artery blood flow measurements, and a clinical evaluation for vascular steal. Temperature measurements were extracted from the thermograms, including bilateral ann and hand (Fistula-Non-Fistula) differences, for comparison with derived AVF blood flow and steal grading. Derived AVF blood flow ranged from 30 to 1950 ml min(-1), with a mean rate close to one litre per minute. Thermography detected the warmer superficial veins in proximity to the patent fistulas, with bilateral differences in fistula region skin temperature correlated with derived AVF blood flow (using maximum temperature measurements the correlation was +0.71 [p < 0.01]; and using mean temperature measurements the correlation was +0.56 [p < 0.05]). When thermography measurements were compared with the clinical assessment of steal the mean hand temperature differences separated steal from non-steal patients with an accuracy of greater than 90%. In summary, we have now demonstrated the potential clinical value of medical infrared thermography for assessing AVF function in renal patients.
引用
收藏
页码:51 / 60
页数:10
相关论文
共 25 条
[1]  
ALLEN J, 2004, P 10 ANN SCI M I PHY
[2]  
Ammer K., 1995, THERMAL IMAGE MED BI
[3]   Contrast-enhanced MR angiography in the assessment of arteriovenous fistula after renal transplant biopsy [J].
Bagga, H ;
Bis, KG .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1999, 172 (06) :1509-1511
[4]  
Besarab A, 1997, ASAIO J, V43, pM539
[5]  
BLAND JM, 1995, INTRO MED STAT
[6]  
DEPNER TA, 1995, ASAIO J, V41, P749
[7]   Accuracy of Doppler ultrasound in diagnosing anatomic stenosis of hemodialysis arteriovenous access as compared with fistulography [J].
Gadallah, MF ;
Paulson, WD ;
Vickers, B ;
Work, J .
AMERICAN JOURNAL OF KIDNEY DISEASES, 1998, 32 (02) :273-277
[8]   USE OF THE FISTULA ASSESSMENT MONITOR TO DETECT STENOSES IN ACCESS FISTULAS [J].
GANI, JS ;
FOWLER, PR ;
STEINBERG, AW ;
WLODARCZYK, JH ;
NANRA, RS ;
HIBBERD, AD .
AMERICAN JOURNAL OF KIDNEY DISEASES, 1991, 17 (03) :303-306
[9]   Steal syndrome complicating hemodialysis access procedures: Can it be predicted? [J].
Goff, CD ;
Sato, DT ;
Bloch, PHS ;
DeMasi, RJ ;
Gregory, RT ;
Gayle, RG ;
Parent, FN ;
Meier, GH ;
Wheeler, JR .
ANNALS OF VASCULAR SURGERY, 2000, 14 (02) :138-144
[10]  
Jung A., 1998, THERMOGRAPHIC METHOD