Evaluation of the skin sensitizing potential of biodegradable magnesium alloys

被引:81
作者
Witte, Frank [1 ]
Abeln, Inken [1 ]
Switzer, Elinor [2 ]
Kaese, Volker [3 ]
Meyer-Lindenberg, Andrea [2 ]
Windhagen, Henning [1 ]
机构
[1] Hannover Med Sch, Dept Orthopaed Surg, D-3000 Hannover, Germany
[2] Vet Sch Hannover, Clin Small Domest Anim, Hannover, Germany
[3] Leibniz Univ Hannover, Ctr Biomed Engn, Hannover, Germany
关键词
biodegradable; magnesium; implant; contact dermatitis; sensitizing potential; bone;
D O I
10.1002/jbm.a.31713
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Corroding metals made of magnesium alloys represent a new class of degradable implants for musculoskeletal surgery. These implants may be associated with skin sensitizing reactions because of the release of metal ions. This study was conducted to compare the sensitizing potential of four different magnesium alloys (AZ31, AZ91, WE43, and LAE442) to current implant materials such as titanium (TiAl6V4) and a degradable polymer (SR-PLA96). Solutions and solid chips of these materials were prepared and tested in 156 guinea pigs according to the Magnusson-Kligman test. A standard allergen (hydroxy-cinnamon-aldehyde) causing allergic erythema was used as positive control and a standard irritant (sodium-lauryl-sulfate) causing local skin irritation for less than 24 h was used as negative control. All erythema were graded immediately and 24 h after patch removal by three independent observers. Histomorphological analyses were performed on skin biopsies taken 24 h after patch removal. We found that initial erythema in animals treated with solid chips diminished within 24 h and were caused by local skin irritation. Local skin irritation was also determined in erythema remaining for 24 h after patch removal in animals treated with dissolved test materials. No allergenic reactions according to the histomorphological criteria were observed in skin biopsies. We conclude that no skin sensitizing potential were detected for standard materials as well as for all tested magnesium alloys by the used methods. (C) 2007 Wiley Periodicals, Inc.
引用
收藏
页码:1041 / 1047
页数:7
相关论文
共 32 条
[1]
Contact sensitivity to aluminum [J].
Akyol, A ;
Boyvat, A ;
Kundakçi, N .
INTERNATIONAL JOURNAL OF DERMATOLOGY, 2004, 43 (12) :942-943
[2]
SUPPRESSION OF THE ALLERGIC CONTACT REACTION IN THE GUINEA-PIG BY CYCLOSPORIN-A [J].
ANDERSON, C ;
GROTH, O .
INTERNATIONAL ARCHIVES OF ALLERGY AND APPLIED IMMUNOLOGY, 1985, 78 (04) :396-400
[3]
ANIMAL-MODEL FOR ASSESSMENT OF SKIN IRRITANCY [J].
ANDERSON, C ;
SUNDBERG, K ;
GROTH, O .
CONTACT DERMATITIS, 1986, 15 (03) :143-151
[4]
ANDERSSON L, 1985, ANIM BLOOD GROUPS BI, V16, P1
[6]
LITHIUM ALUMINUM-ALLOYS - A PROBLEM MATERIAL FOR BIOLOGICAL MONITORING [J].
BENCZE, K ;
PELIKAN, C ;
BAHEMANNHOFFMEISTER, A ;
KRONSEDER, A .
SCIENCE OF THE TOTAL ENVIRONMENT, 1991, 101 (1-2) :83-90
[7]
CLAES L E, 1992, Clinical Materials, V10, P41, DOI 10.1016/0267-6605(92)90083-6
[8]
Epidermal cytokines in experimental contact dermatitis [J].
Corsini, E ;
Galli, CL .
TOXICOLOGY, 2000, 142 (03) :203-211
[9]
CUTANEOUS BASOPHIL HYPERSENSITIVITY .2. A LIGHT AND ELECTRON MICROSCOPIC DESCRIPTION [J].
DVORAK, HF ;
DVORAK, AM ;
SIMPSON, BA ;
RICHERSON, HB ;
LESKOWITZ, S ;
KARNOVSKY, MJ .
JOURNAL OF EXPERIMENTAL MEDICINE, 1970, 132 (03) :558-+
[10]
GROTH O, 1979, ACTA DERM-VENEREOL, V59, P129