Radiation and chemical crosslinking promote strain hardening behavior and molecular alignment in ultra high molecular weight polyethylene during multi-axial loading conditions

被引:79
作者
Kurtz, SM
Pruitt, LA
Jewett, CW
Foulds, JR
Edidin, AA
机构
[1] Exponent Failure Anal Associates Inc, Philadelphia, PA 19103 USA
[2] Univ Calif Berkeley, Dept Mech Engn, Berkeley, CA 94720 USA
[3] Stryker Howmedica Osteon, Allendale, NJ 07410 USA
关键词
UHMWPE; polyethylene; crosslinking; peroxide; irradiation;
D O I
10.1016/S0142-9612(99)00038-1
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The mechanical behavior and evolution of crystalline morphology during large deformation of eight types of virgin and crosslinked ultra high molecular weight polyethylene (UHMWPE) were studied using the small punch test and transmission electron microscopy (TEM). We investigated the hypothesis that both radiation and chemical crosslinking hinder molecular mobility at large deformations, and hence promote strain hardening and molecular alignment during the multiaxial loading of the small punch test. Chemical crosslinking of UHMWPE was performed using 0.25% dicumyl peroxide (GHR 8110, GUR 1020 and 1050), and radiation crosslinking was performed using 150 kGy of electron beam radiation (GUR 1150). Crosslinking increased the ultimate load at failure and decreased the ultimate displacement of the polyethylenes during the small punch test. Crosslinking also increased the near-ultimate hardening behavior of the polyethylenes. Transmission electron microscopy was used to characterize the crystalline morphology of the bulk material, undeformed regions of the small punch test specimens, and deformed regions of the specimens oriented perpendicular and parallel to the punch direction. In contrast with the virgin polyethylenes, which showed only subtle evidence of lamellar alignment, the crosslinked polyethylenes exhibited enhanced crystalline lamellae orientation after the small punch test, predominantly in the direction parallel to the punch direction or deformation axis. Thus, the results of this study support the hypothesis that crosslinking promotes strain hardening during multiaxial loading because of increased resistance to molecular mobility at large deformations effected by molecular alignment. The data also illustrate the sensitivity of large deformation mechanical behavior and crystalline morphology to the method of crosslinking and resin of polyethylene, (C) 1999 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:1449 / 1462
页数:14
相关论文
共 35 条
[1]   PLASTIC-FLOW OF ISOTROPIC POLYMERS [J].
BOWDEN, PB ;
JUKES, JA .
JOURNAL OF MATERIALS SCIENCE, 1972, 7 (01) :52-&
[2]   EFFECTS OF POLYETHYLENE MOLECULAR-STRUCTURE ON PEROXIDE CROSS-LINKING OF LOW-DENSITY POLYETHYLENE [J].
BREMNER, T ;
RUDIN, A ;
HARIDOSS, S .
POLYMER ENGINEERING AND SCIENCE, 1992, 32 (14) :939-943
[3]   PEROXIDE MODIFICATION OF LINEAR LOW-DENSITY POLYETHYLENE - A COMPARISON OF DIALKYL PEROXIDES [J].
BREMNER, T ;
RUDIN, A .
JOURNAL OF APPLIED POLYMER SCIENCE, 1993, 49 (05) :785-798
[5]   MACROSCOPIC AND MICROSCOPIC WEAR MECHANISMS IN ULTRA-HIGH-MOLECULAR-WEIGHT POLYETHYLENE [J].
COOPER, JR ;
DOWSON, D ;
FISHER, J .
WEAR, 1993, 162 :378-384
[6]  
EDIDIN AA, 1999, IN PRESS J ARTHROPLA
[7]  
FOULDS JR, 1995, J TEST EVAL, V23, P3
[8]  
Gul R., 1997, THESIS MIT BOSTON
[9]   Wear of polyethylene acetabular components in total hip arthroplasty - An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations [J].
Jasty, M ;
Goetz, DD ;
Bragdon, CR ;
Lee, KR ;
Hanson, AE ;
Elder, JR ;
Harris, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (03) :349-358
[10]  
JASTY M, 1997, ORTHOP RES SOC, V43, P785