Respiration-induced B0 field fluctuation compensation in balanced SSFP:: Real-time approach for transition-band SSFP fMRI

被引:38
作者
Lee, J
Santos, JM
Conolly, SM
Miller, KL
Hargreaves, BA
Pauly, JM
机构
[1] Stanford Univ, Dept Elect Engn, Magnet Resonance Syst Res Lab, Stanford, CA 94305 USA
[2] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[3] Univ Calif Berkeley, Dept Bioengn, Berkeley, CA 94720 USA
[4] Univ Oxford, FMRIB, Oxford, England
关键词
fMRI; steady state; real time; respiration; SSFP;
D O I
10.1002/mrm.20879
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In functional MRI (fMRI) the resonance frequency shift induced from respiration is a major source of physiological noise. In transition-band SSFP fMRI the respiration-induced resonance offset not only increases noise interference, it also shifts the activation band. This leads to a reduction in the contrast-to-noise ratio (CNR) and the potential for varying contrast levels during the experiment. A novel real-time method that compensates for the respiration-induced resonance offset frequency is presented. This method utilizes free induction decay (FID) phase information to measure the resonance offset. For compensation, one can update the resonant frequency in real time by changing the transmit RF pulse and receiver phases to track the measured offset. The results show decreased signal power in the respiration frequency band and increased numbers of activated voxels with higher Z-scores compared to uncompensated experiments.
引用
收藏
页码:1197 / 1201
页数:5
相关论文
共 21 条
[1]   Analysis and compensation of eddy currents in balanced SSFP [J].
Bieri, O ;
Markl, M ;
Scheffler, K .
MAGNETIC RESONANCE IN MEDICINE, 2005, 54 (01) :129-137
[2]   Reduction of physiological fluctuations in fMRI using digital filters [J].
Biswal, B ;
DeYoe, EA ;
Hyde, JS .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (01) :107-113
[3]   Noise suppression digital filter for functional magnetic resonance imaging based on image reference data [J].
Buonocore, MH ;
Maddock, RJ .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (03) :456-469
[4]  
CHO ZH, 1995, P 3 ANN M SMR NIC FR, P806
[5]   IMPACT: Image-based physiological artifacts estimation and correction technique for functional MRI [J].
Chuang, KH ;
Chen, JH .
MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (02) :344-353
[6]   Frequency-modulated steady-state free precession imaging [J].
Foxall, DL .
MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (03) :502-508
[7]   PHASE AND INTENSITY ANOMALIES IN FOURIER TRANSFORM NMR [J].
FREEMAN, R ;
HILL, HDW .
JOURNAL OF MAGNETIC RESONANCE, 1971, 4 (03) :366-&
[8]  
Glover GH, 2000, MAGNET RESON MED, V44, P162, DOI 10.1002/1522-2594(200007)44:1<162::AID-MRM23>3.0.CO
[9]  
2-E
[10]   IMAGE-FORMATION BY NUCLEAR MAGNETIC-RESONANCE - SENSITIVE-POINT METHOD [J].
HINSHAW, WS .
JOURNAL OF APPLIED PHYSICS, 1976, 47 (08) :3709-3721