Improved coronary disease detection with quantitative attenuation-corrected Tl-201 images

被引:47
作者
Shotwell, M
Singh, BM
Fortman, C
Bauman, BD
Lukes, J
Gerson, MC
机构
[1] Univ Cincinnati, Dept Internal Med, Div Cardiol, Cincinnati, OH 45267 USA
[2] Univ Cincinnati, Dept Radiol, EL Saenger Radioisotope Lab, Cincinnati, OH 45267 USA
关键词
thallium; 201; attenuation correction; scatter correction; coronary artery disease;
D O I
10.1067/mnc.2002.119252
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. The normal distribution of myocardial tracer activity is different in attenuation-corrected images compared with uncorrected images. We therefore postulated that quantitation of attenuation-corrected thallium 201 images with direct comparison to a database of healthy subjects could improve detection of coronary artery stenoses. Methods and Results. In 49 patients with angiographic evidence of coronary artery disease and 69 patients with a less than 5% likelihood of coronary artery disease, tomographic Tl-201 myocardial imaging was performed by means of a triple-headed camera with fan-beam collimators and the images were processed with attenuation correction, with attenuation and Compton scatter correction, and without correction. Images were subjectively interpreted in a blinded manner. From the group of 69 patients with a low likelihood of coronary disease, scintigraphic data from the first 20 men and the first 20 women were used to generate normal reference ranges for each of the 3 image-processing methods. Data from the 49 patients with angiographic coronary disease and from the remaining 29 patients with a low likelihood of coronary disease were then analyzed quantitatively in comparison to the gender-matched normal databases. With visual interpretation, attenuation-corrected images yielded an improved detection rate compared with uncorrected images (79% vs 46%, P=.008) for 70% or greater left anterior descending coronary artery stenoses. Attenuation-corrected images also provided an improved normalcy rate for the right coronary artery territory (91% vs 75%, P=.006) and an improved normalcy rate when the images were analyzed overall by patient (88% vs 74%, P=.009). With quantitative analysis, attenuation correction yielded an improved detection rate for patients with a significant coronary artery stenosis compared with uncorrected images (90% vs 71%, P=.016). Defect extent was also increased by attenuation correction. Compared with attenuation correction alone, the addition of scatter correction yielded no incremental improvement in detection rate or normalcy rate. Attenuation-scatter correction did produce a nonsignificant trend toward detection of more of the total number of stenotic arteries compared with uncorrected images (60 vs 48 of 79 arteries). Conclusions. With the use of a triple-headed camera with fan-beam collimation, visual analysis of attenuation-corrected Tl-201 images improved detection of left anterior descending coronary artery disease without loss of detection in other coronary territories. Furthermore, quantitative analysis of attenuation-corrected Tl-201 images improved the overall detection rate for coronary artery disease in patients without compromise of the normalcy rate.
引用
收藏
页码:52 / 61
页数:10
相关论文
共 23 条
[1]   MAXIMAL OXYGEN INTAKE AND NOMOGRAPHIC ASSESSMENT OF FUNCTIONAL AEROBIC IMPAIRMENT IN CARDIOVASCULAR DISEASE [J].
BRUCE, RA ;
KUSUMI, F ;
HOSMER, D .
AMERICAN HEART JOURNAL, 1973, 85 (04) :546-562
[2]   Evaluation of an attenuation correction method for thallium-201 myocardial perfusion tomographic imaging of patients with low likelihood of coronary artery disease [J].
Chouraqui, P ;
Livschitz, S ;
Sharir, T ;
Wainer, N ;
Wilk, M ;
Moalem, I ;
Baron, J .
JOURNAL OF NUCLEAR CARDIOLOGY, 1998, 5 (04) :369-377
[3]   TL-201 SINGLE-PHOTON EMISSION COMPUTED-TOMOGRAPHY WITH INTRAVENOUS DIPYRIDAMOLE TO DIAGNOSE CORONARY-ARTERY DISEASE [J].
DEPUEY, EG ;
GUERTLERKRAWCZYNSKA, E ;
DAMATO, PH ;
PATTERSON, RE .
CORONARY ARTERY DISEASE, 1990, 1 (01) :75-82
[4]   ANALYSIS OF PROBABILITY AS AN AID IN THE CLINICAL-DIAGNOSIS OF CORONARY-ARTERY DISEASE [J].
DIAMOND, GA ;
FORRESTER, JS .
NEW ENGLAND JOURNAL OF MEDICINE, 1979, 300 (24) :1350-1358
[5]  
FICARO EP, 1995, J NUCL MED, V36, P921
[6]   Simultaneous transmission emission myocardial perfusion tomography diagnostic accuracy of attenuatian-corrected Tc-99m-sestamibi single-photon emission computed tomography [J].
Ficaro, EP ;
Fessler, JA ;
Shreve, PD ;
Kritzman, JN ;
Rose, PA ;
Corbett, JR .
CIRCULATION, 1996, 93 (03) :463-473
[7]   Attenuation-corrected thallium-201 single-photon emission tomography using a gadolinium-153 moving line source: clinical value and the impact of attenuation correction on the extent and severity of perfusion abnormalities [J].
Gallowitsch, HJ ;
Sykora, J ;
Mikosch, P ;
Kresnik, E ;
Unterweger, O ;
Molnar, M ;
Grimm, G ;
Lind, P .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1998, 25 (03) :220-228
[8]  
GARCIA EV, 1985, J NUCL MED, V26, P17
[9]   Multicenter clinical trial to evaluate the efficacy of correction for photon attenuation and scatter in SPECT myocardial perfusion imaging [J].
Hendel, RC ;
Berman, DS ;
Cullom, SJ ;
Follansbee, W ;
Heller, GV ;
Kiat, H ;
Groch, MW ;
Mahmarian, JJ .
CIRCULATION, 1999, 99 (21) :2742-2749
[10]   COMPARISON OF DIPYRIDAMOLE-HANDGRIP TEST AND BICYCLE EXERCISE TEST FOR THALLIUM TOMOGRAPHIC IMAGING [J].
HUIKURI, HV ;
KORHONEN, UR ;
AIRAKSINEN, KEJ ;
IKAHEIMO, MJ ;
HEIKKILA, J ;
TAKKUNEN, JT .
AMERICAN JOURNAL OF CARDIOLOGY, 1988, 61 (04) :264-268