Biomedical implants and devices: Assessment of magnetic field interactions with a 3.0-Tesla MR system

被引:154
作者
Shellock, FG
机构
[1] Univ So Calif, Keck Sch Med, Los Angeles, CA USA
[2] Inst Magnet Resonance Safety Educ & Res, Los Angeles, CA USA
关键词
magnetic resonance; safety; magnetic resonance imaging; implants; bioeffects; high field;
D O I
10.1002/jmri.10207
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Purpose: To evaluate magnetic field interactions for 109 different biomedical implants and devices in association with exposure to a 3.0-Tesla magnetic resonance (MR) system. Materials and Methods: A total of 109 implants and devices (aneurysm clips, 32: clips, fasteners, and staples, 10; coils and stents, 10; heart valve prostheses and annuloplasty rings, 12; orthopedic implants, five; suture materials, 13; vascular access ports and accessories, 13; miscellaneous implants and devices, 14) were tested for magnetic field interactions at 3.0-Tesla using previously-described, standardized techniques to assess magnetic field translational attraction and torque. Results: The deflection angles and torque measurements ranged, respectively, from 0 to 16degrees and 0 to +2 for the aneurysm clips; 0 to 90degrees and 0 to +4 for the clips, fasteners, and staples; 0 to 47degrees and 0 to +4 for the coils and stents; 0 to 4degrees and 0 to + 1 for the heart valve prostheses and annuloplasty rings; 0 to 12degrees and 0 to +2 for the orthopedic implants; 0 to 13degrees and 0 to + 2 for the suture materials; 0 to 52degrees and 0 to + 4 for the vascular access ports and accessories; and 0 to 28degrees and 0 to +3 for the miscellaneous implants and devices. Conclusion: Of the 109 implants and devices assessed for magnetic field interactions at 3.0-Tesla, four (4%) are potentially unsafe based on deflection angle criteria. The implications of these results for patients undergoing MR procedures at 3.0-Tesla is discussed. Notably, these results are specific to the 3.0-Tesla MR system used for this evaluation.
引用
收藏
页码:721 / 732
页数:12
相关论文
共 57 条
[1]
[Anonymous], 2002, ANN BOOK ASTM STAND, V13.01, P1576
[2]
[Anonymous], EUR RADIOL
[3]
BARRAFATO D, 1984, CAN J SURG, V27, P509
[4]
MAGNETIC-FIELD EFFECTS ON SURGICAL LIGATION CLIPS [J].
BROWN, MA ;
CARDEN, JA ;
COLEMAN, RE ;
MCKINNEY, R ;
SPICER, LD .
MAGNETIC RESONANCE IMAGING, 1987, 5 (06) :443-453
[5]
POTENTIAL HAZARDS IN NMR IMAGING - HEATING EFFECTS OF CHANGING MAGNETIC-FIELDS AND RF FIELDS ON SMALL METALLIC IMPLANTS [J].
DAVIS, PL ;
CROOKS, L ;
ARAKAWA, M ;
MCREE, R ;
KAUFMAN, L ;
MARGULIS, AR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1981, 137 (04) :857-860
[6]
Edwards M B, 2000, J Magn Reson Imaging, V12, P363, DOI 10.1002/1522-2586(200008)12:2<363::AID-JMRI21>3.0.CO
[7]
2-3
[8]
WALLSTENT METALLIC BILIARY ENDOPROSTHESIS - MR IMAGING CHARACTERISTICS [J].
GIRARD, MJ ;
HAHN, PF ;
SAINI, S ;
DAWSON, SL ;
GOLDBERG, MA ;
MUELLER, PR .
RADIOLOGY, 1992, 184 (03) :874-876
[9]
Magnetic resonance safety testing of a newly-developed fiber-optic cardiac pacing lead [J].
Greatbatch, W ;
Miller, V ;
Shellock, FG .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2002, 16 (01) :97-103
[10]
HASSLER M, 1986, Journal de Radiologie (Paris), V67, P661