Leakage of fluid around low-pressure tracheal tube cuffs

被引:35
作者
Oikkonen, M [1 ]
Aromaa, U [1 ]
机构
[1] HELSINKI UNIV HOSP, MEILAHTI HOSP, DEPT ANAESTHESIA, HELSINKI 00290, FINLAND
关键词
equipment; tubes; tracheal; lung; aspiration;
D O I
10.1111/j.1365-2044.1997.149-az0153.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The aim of the study was to evaluate leakage of liquid past the low-pressure cuffs of tracheal tubes. Ten samples of each of nine different types of tubes were tested in a PVC mock trachea, using intracuff pressures of 20, 30, 40 and 50 cmH(2)O. In five types of tubes, 6-10 cuffs allowed profuse leakage (> 20 ml water in 5 min) even at the highest intracuff pressure, i.e. 50 cmH(2)O. In the most efficient tube, all the cuffs were leak-proof (leakage < 5 ml in 5 min) at 40 cmH(2)O and in the second best type the cuffs were leak-proof at 50 cmH(2)O. The leakage of fluid past the tracheal tube remains an unresolved problem with low-pressure cuffs.
引用
收藏
页码:567 / 569
页数:3
相关论文
共 6 条
[1]   INTRACUFF PRESSURES IN ENDOTRACHEAL AND TRACHEOSTOMY TUBES - RELATED CUFF PHYSICAL CHARACTERISTICS [J].
BERNHARD, WN ;
YOST, L ;
JOYNES, D ;
COTHALIS, S ;
TURNDORF, H .
CHEST, 1985, 87 (06) :720-725
[2]   ADJUSTMENT OF INTRACUFF PRESSURE TO PREVENT ASPIRATION [J].
BERNHARD, WN ;
COTTRELL, JE ;
SIVAKUMARAN, C ;
PATEL, K ;
YOST, L ;
TURNDORF, H .
ANESTHESIOLOGY, 1979, 50 (04) :363-366
[3]   RATIONALIZATION OF THE SELECTION OF TRACHEAL TUBES [J].
CHANDLER, M ;
CRAWLEY, BE .
BRITISH JOURNAL OF ANAESTHESIA, 1986, 58 (01) :111-116
[4]   FAILURE OF A HIGH-COMPLIANCE LOW-PRESSURE CUFF TO PREVENT ASPIRATION [J].
PAVLIN, EG ;
VANNIMWEGEN, D ;
HORNBEIN, TF .
ANESTHESIOLOGY, 1975, 42 (02) :216-219
[5]  
PETRING OU, 1986, ANESTH ANALG, V65, P777
[6]   ENDOTRACHEAL-TUBE CUFF RESIDUAL VOLUME AND LATERAL WALL PRESSURE IN A MODEL TRACHEA [J].
TONNESEN, AS ;
VEREEN, L ;
ARENS, JF .
ANESTHESIOLOGY, 1981, 55 (06) :680-683