Significant Change or Loss of Intraoperative Monitoring Data A 25-Year Experience in 12,375 Spinal Surgeries

被引:77
作者
Raynor, Barry L. [1 ]
Bright, Joseph D. [1 ]
Lenke, Lawrence G. [2 ]
Rahman, Ra'Kerry K.
Bridwell, Keith H.
Riew, K. Daniel
Buchowski, Jacob M.
Luhmann, Scott J.
Padberg, Anne M. [1 ]
机构
[1] Barnes Jewish Hosp, St Louis, MO 63110 USA
[2] Washington Univ, Sch Med, Dept Orthopaed Surg, St Louis, MO 63110 USA
关键词
spinal surgery; revision spinal surgery; intraoperative monitoring; neurological deficit; MOTOR-EVOKED-POTENTIALS; SCOLIOSIS SURGERY; EFFICACY;
D O I
10.1097/BRS.0b013e31827aafb9
中图分类号
R74 [神经病学与精神病学];
学科分类号
100204 [神经病学];
摘要
Study Design. Retrospective. Objective. The purpose of this study was to report the spectrum of intraoperative events responsible for a loss or significant change in intraoperative monitoring (IOM) data. Summary of Background Data. The efficacy of spinal cord/nerve root monitoring is demonstrated in a large, single institution series of patients, involving all levels of the spinal column (occiput to sacrum) and all spinal surgical procedures. Methods. Multimodality IOM included somatosensory-evoked potentials, descending neurogenic-evoked potentials, neurogenic motor-evoked potentials, and spontaneous and triggered electromyography. A total of 12,375 patients who underwent surgery for spinal pathology between January 1985 and December 2010 were reviewed. There were 59.3% female patients (7178) and 40.7% male patients (5197). Procedures by spinal level were as follows: cervical 29.7% (3671), thoracic/thoracolumbar 45.4% (5624), and lumbosacral 24.9% (3080). Age at the time of surgery was as follows: older than 18 years, 72.7% (242/8993) and younger than 18 years, 27.3% (144/3382). A total of 77.8% (9633) patients underwent primary surgical procedures and 22.2% (2742) patients underwent revision surgical procedures. Results. A total of 406 instances of IOM data change/loss occurred in 386 of 12,375 (3.1%) patients. Causes for data degradation/loss included the following: instrumentation (n = 131), positioning (n = 85), correction (n = 56), systemic (n = 49), unknown (n = 24), and focal spinal cord compression (n = 15). Data loss/change was seen in revision (6.1%/167 patients) surgical procedures more commonly than in primary procedures (2.3%/219 patients; P < 0.0001). Data improvement was demonstrated by 88.7% (n = 360) after intervention versus 11.3% (n = 46) with no improvement in IOM data. One patient with improved data after intervention versus 14 with no improvement despite intervention had a permanent neurological deficit (P < 0.0001). Conclusion. IOM data identified 386 (3.1%) patients with loss/degradation of data in 12,375 spinal surgical procedures. Fortunately, in 93.3% of patients, intervention led to data recovery and no neurological deficits. Reduction from a potential (worst-case scenario) 3.1% (386) of patients with significant change/loss of IOM data to a permanent neurological deficit rate of 0.12% (15) patients was achieved (P < 0.0001), thus confirming efficacy of IOM.
引用
收藏
页码:E101 / E108
页数:8
相关论文
共 27 条
[1]
[Anonymous], 2009, SCOLIOSIS RESEARCH S
[2]
ASSESSMENT OF CORTICOSPINAL AND SOMATOSENSORY CONDUCTION SIMULTANEOUSLY DURING SCOLIOSIS SURGERY [J].
BURKE, D ;
HICKS, R ;
STEPHEN, J ;
WOODFORTH, I ;
CRAWFORD, M .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1992, 85 (06) :388-396
[3]
Alarm criteria for motor-evoked potentials - What's wrong with the "Presence-or-Absence" approach? [J].
Calancie, Blair ;
Molano, Maria R. .
SPINE, 2008, 33 (04) :406-414
[4]
DAWSON EG, 1991, SPINE, V16, pS361
[5]
Deletis V., 2002, Neurophysiology in neurosurgery a modern intraoperative approach, P25, DOI [10.1016/B978-012209036-3/50004-4, DOI 10.1016/B978-012209036-3/50004-4]
[6]
Neural complications in the surgical treatment of adolescent idiopathic scoliosis [J].
Diab, Mohammad ;
Smith, Amanda R. ;
Kuklo, Timothy R. .
SPINE, 2007, 32 (24) :2759-2763
[7]
Ecker ML, 1996, J SPINAL DISORD, V9, P159
[8]
The Evidence for Intraoperative Neurophysiological Monitoring in Spine Surgery Does It Make a Difference? [J].
Fehlings, Michael G. ;
Brodke, Darrel S. ;
Norvell, Daniel C. ;
Dettori, Joseph R. .
SPINE, 2010, 35 (09) :S37-S46
[9]
Comparison of transcranial electric motor and somatosensory evoked potential monitoring during cervical spine surgery [J].
Hilibrand, AS ;
Schwartz, DM ;
Sethuraman, V ;
Vaccaro, AR ;
Albert, TJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (06) :1248-1253
[10]
TRIGGERED ELECTROMYOGRAPHIC THRESHOLD FOR ACCURACY OF PEDICLE SCREW PLACEMENT - AN ANIMAL-MODEL AND CLINICAL CORRELATION [J].
LENKE, LG ;
PADBERG, AM ;
RUSSO, MH ;
BRIDWELL, KH ;
GELB, DE .
SPINE, 1995, 20 (14) :1585-1591