Trace level voltammetric determination of manganese, iron and chromium in real samples in the presence of each other

被引:13
作者
Locatelli, C
机构
[1] Department of Chemistry G. Ciamician, University of Bologna, I-40126 Bologna
关键词
voltammetry; manganese; iron; chromium; reference materials;
D O I
10.1016/0039-9140(95)01703-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The determination of manganese in the presence of iron and chromium by differential pulse voltammetry and fundamental harmonic alternating current voltammetry was compared, including the case of very high element concentration ratios. The voltammetric measurements were carried out using a stationary mercury electrode in ammonia-ammonium chloride buffer (pH 9.6). The analytical procedure was verified by the analysis of the standard reference materials Portland Cement BCS 372, Spectrographic Zinc Spelter NBS-SRM 631, Stainless Steel(AISI 321) NBS-SRM 121d and Highly Alloyed Steel Eurostandard 281-1. Precision and accuracy, expressed as relative standard deviation and relative error respectively, were of the order of 3-5%,, while the detection limit for each element was around 1 x 10(-9) M. The standard addition technique improved the resolution of the voltammetric methods, within a maximum experimental error of 5%, even in the case of very high concentration ratios, that is outside the non-interference concentration ratios 69:1 > c(Fe):c(Mn) > 1:74; 35:1 > c(Fe):c(Cr) > 1:30 and 63:1 > c(Fe):c(Mn) > 1:65; 32:1 > c(Fe):c(Cr) > 1:31 for the differential pulse and alternating current techniques respectively, extrapolating the linear section of the i(p) vs. concentration analytical calibration function for the element present at the lowest concentration, In contrast, the element with the greatest concentration was determined by the relevant calibration curve.
引用
收藏
页码:45 / 54
页数:10
相关论文
共 24 条
[1]   DETERMINATION OF MANGANESE IN WATER SAMPLES BY GALVANOSTATIC STRIPPING CHRONOPOTENTIOMETRY IN A FLOW-THROUGH CELL [J].
BEINROHR, E ;
CSEMI, P ;
ROJAS, FJ ;
HOFBAUEROVA, H .
ANALYST, 1994, 119 (06) :1355-1359
[2]  
BETTI M, 1992, ANN CHIM-ROME, V82, P339
[3]  
Bond A.M., 1980, MODERN POLAROGRAPHIC
[4]   REDUCTIVE POTENTIOMETRIC STRIPPING ANALYSIS [J].
CHRISTENSEN, JK ;
KRYGER, L .
ANALYTICA CHIMICA ACTA, 1980, 118 (01) :53-64
[5]   DETERMINATION OF MANGANESE AT NG ML LEVELS IN NATURAL-WATERS BY DIFFERENTIAL PULSE POLAROGRAPHY [J].
COLOMBINI, MP ;
FUOCO, R .
TALANTA, 1983, 30 (12) :901-905
[6]   POTENTIOMETRIC STRIPPING ANALYSIS FOR MANGANESE(II) IN NATURAL-WATERS [J].
ESKILSSON, H ;
TURNER, DR .
ANALYTICA CHIMICA ACTA, 1984, 161 (JUL) :293-302
[7]   APPLICATION OF 2ND HARMONIC ALTERNATING-CURRENT ANODIC-STRIPPING VOLTAMMETRY IN THE SIMULTANEOUS TRACE AND ULTRA TRACE-METAL DETERMINATION OF A MULTICOMPONENT SYSTEM [J].
FAGIOLI, F ;
LOCATELLI, C ;
LANDI, S ;
VECCHIETTI, R .
ELECTROANALYSIS, 1989, 1 (05) :449-455
[8]  
FAGIOLI F, 1974, ANN CHIM-ROME, V64, P633
[9]  
HALD A, 1952, STATISTICAL THEORY E
[10]   DETERMINATION OF MN(II) IN TAP WATER [J].
KALAVSKA, D .
INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 1991, 45 (03) :159-167