Range modulation in proton therapy - an optimization technique for clinical and experimental applications

被引:14
作者
Gardey, KU
Oelfke, U
Lam, GKY
机构
[1] Humboldt Univ, Fak Med, Virchow Klinikum, D-13353 Berlin, Germany
[2] Deutsch Krebsforschungszentrum, Abt Med Phys, D-69120 Heidelberg, Germany
[3] Ottawa Reg Canc Ctr, Ottawa, ON K1Y 4K7, Canada
关键词
D O I
10.1088/0031-9155/44/6/402
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A fast optimization algorithm for range modulation in clinical and experimental applications of proton therapy is described. The method is versatile towards the number of parameters provided for range modulation, i.e. the trade-off between accuracy and simplicity of the latter can be chosen freely. The approach is, therefore, adaptable to most operating proton therapy facilities. It requires only a few basic measurements as input data and results in a depth dose uniformity of better than 2%. A typical calculation takes less than 90 s on a DEC VAXstation 3100, FORTRAN. The method has been extensively tested at the TRIUMF Proton Therapy Facility in Vancouver, BC.
引用
收藏
页码:N81 / N88
页数:8
相关论文
共 4 条
[1]   A PROTOTYPE BEAM DELIVERY SYSTEM FOR THE PROTON MEDICAL ACCELERATOR AT LOMA-LINDA [J].
COUTRAKON, G ;
BAUMAN, M ;
LESYNA, D ;
MILLER, D ;
NUSBAUM, J ;
SLATER, J ;
JOHANNING, J ;
MIRANDA, J ;
DELUCA, PM ;
SIEBERS, J ;
LUDEWIGT, B .
MEDICAL PHYSICS, 1991, 18 (06) :1093-1099
[2]   A GENERAL-SOLUTION TO CHARGED-PARTICLE BEAM FLATTENING USING AN OPTIMIZED DUAL-SCATTERING-FOIL TECHNIQUE, WITH APPLICATION TO PROTON THERAPY BEAMS [J].
GRUSELL, E ;
MONTELIUS, A ;
BRAHME, A ;
RIKNER, G ;
RUSSELL, K .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (12) :2201-2216
[3]   RANGE MODULATORS FOR PROTONS AND HEAVY-IONS [J].
KOEHLER, AM ;
SCHNEIDER, RJ ;
SISTERSON, JM .
NUCLEAR INSTRUMENTS & METHODS, 1975, 131 (03) :437-440
[4]  
WOUTERS BG, 1996, RADIAT RES, V146, P153