Accurate three-dimensional reconstruction of intravascular ultrasound data - Spatially correct three-dimensional reconstructions

被引:72
作者
Evans, JL
Ng, KH
Wiet, SG
Vonesh, MJ
Burns, WB
Radvany, MG
Kane, BJ
Davidson, CJ
Roth, SI
Kramer, BL
Meyers, SN
McPherson, DD
机构
[1] NORTHWESTERN UNIV,SCH MED,DEPT MED,CHICAGO,IL 60611
[2] NORTHWESTERN UNIV,SCH MED,DEPT PATHOL,CHICAGO,IL 60611
关键词
ultrasonics; coronary disease; imaging;
D O I
10.1161/01.CIR.93.3.567
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background The geometrical accuracy of conventional three-dimensional (3D) reconstruction methods for intravascular ultrasound (IVUS) data (coronary and peripheral) is hampered by the inability to register spatial image orientation and by respiratory and cardiac motion. The objective of this work was the development of improved IVUS reconstruction techniques. Methods and Results We developed a 3D position registration method that identifies the spatial coordinates of an in situ IVUS catheter by use of simultaneous EGG-gated biplane digital cinefluoroscopy; To minimize distortion, coordinates underwent pincushion correction and were referenced to a standardized calibration cube. Gated IVUS data were acquired digitally, and the spatial locations of the imaging planes were matched to their corresponding coordinates. Image points were then transformed relative to their respective 3D coordinates, rendered in binary voxel format, resliced, and displayed on an image-processing workstation for off-line analysis. The method was tested by use of phantoms (straight tube, 360 degrees circle, 240 degrees spiral) and an in vitro coronary artery model. In vivo feasibility was assessed in patients who underwent routine interventional coronary procedures accompanied by IVUS evaluation. Actual versus calculated point locations were within 1.0+/-0.3 mm of each other (n=39). Calculated phantom volumes were within 4% of actual volumes. Phantom 3D reconstruction appropriately demonstrated complex morphology. Initial patient evaluation demonstrated method feasibility as well as errors if respiratory and ECG gating were not used. Conclusions These preliminary data support the use of this new method of 3D reconstruction of vascular structures with use of combined vascular ultrasound data and simultaneous EGG-gated biplane cinefluoroscopy.
引用
收藏
页码:567 / 576
页数:10
相关论文
共 18 条
  • [1] INVITRO VALIDATION OF 3-DIMENSIONAL INTRAVASCULAR ULTRASOUND FOR THE EVALUATION OF ARTERIAL INJURY AFTER BALLOON ANGIOPLASTY
    COY, KM
    PARK, JC
    FISHBEIN, MC
    LAAS, T
    DIAMOND, GA
    ADLER, L
    MAURER, G
    SIEGEL, RJ
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 20 (03) : 692 - 700
  • [2] ARTERIAL-WALL CHARACTERISTICS DETERMINED BY INTRAVASCULAR ULTRASOUND IMAGING - AN INVITRO STUDY
    GUSSENHOVEN, EJ
    ESSED, CE
    LANCEE, CT
    MASTIK, F
    FRIETMAN, P
    VANEGMOND, FC
    REIBER, J
    BOSCH, H
    VANURK, H
    ROELANDT, J
    BOM, N
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (04) : 947 - 952
  • [3] Gussenhoven W J, 1989, Int J Card Imaging, V4, P105, DOI 10.1007/BF01745140
  • [4] PERCUTANEOUS INTRAVASCULAR-US AS ADJUNCT TO CATHETER-BASED INTERVENTIONS - PRELIMINARY EXPERIENCE IN PATIENTS WITH PERIPHERAL VASCULAR-DISEASE
    ISNER, JM
    ROSENFIELD, K
    LOSORDO, DW
    KELLY, S
    PALEFSKI, P
    LANGEVIN, RE
    RAZVI, S
    PASTORE, JO
    KOSOWSKY, BD
    [J]. RADIOLOGY, 1990, 175 (01) : 61 - 70
  • [5] KITNEY RI, 1989, INT J CARDIAC IMAG, V4, P177
  • [6] 3D-SURFACE RECONSTRUCTION OF INTRAVASCULAR ULTRASOUND IMAGES USING PERSONAL-COMPUTER HARDWARE AND A MOTORIZED CATHETER CONTROL
    KLEIN, HM
    GUNTHER, RW
    VERLANDE, M
    SCHNEIDER, W
    VORWERK, D
    KELCH, J
    HAMM, M
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1992, 15 (02) : 97 - 101
  • [7] MCKAY C R, 1990, Journal of the American College of Cardiology, V15, p16A
  • [8] MCKAY S, 1982, COMPUT BIOMED RES, V15, P455
  • [9] INTRAVASCULAR ULTRASOUND IMAGING - INVITRO VALIDATION AND PATHOLOGICAL CORRELATION
    NISHIMURA, RA
    EDWARDS, WD
    WARNES, CA
    REEDER, GS
    HOLMES, DR
    TAJIK, AJ
    YOCK, PG
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (01) : 145 - 154
  • [10] APPLICATION OF A NEW PHASED-ARRAY ULTRASOUND IMAGING CATHETER IN THE ASSESSMENT OF VASCULAR DIMENSIONS - INVIVO COMPARISON TO CINEANGIOGRAPHY
    NISSEN, SE
    GRINES, CL
    GURLEY, JC
    SUBLETT, K
    HAYNIE, D
    DIAZ, C
    BOOTH, DC
    DEMARIA, AN
    [J]. CIRCULATION, 1990, 81 (02) : 660 - 666