On-line preconcentration system for flame atomic absorption spectrometry using unloaded polyurethane foam: determination of zinc in waters and biological materials

被引:55
作者
Cassella, RJ [1 ]
Bitencourt, DT
Branco, AG
Ferreira, SLC
de Jesus, DS
de Carvalho, MS
Santelli, RE
机构
[1] Escola Tecn Fed Quim, BR-20270021 Rio De Janeiro, Brazil
[2] Univ Fed Fluminense, Dept Geoquim, BR-24020007 Niteroi, RJ, Brazil
[3] Univ Fed Bahia, Inst Quim, BR-40170290 Salvador, BA, Brazil
[4] CNEN, Inst Engn Nucl, Rio De Janeiro, Brazil
关键词
D O I
10.1039/a904997e
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An on-line procedure for the preconcentration and determination of zinc in waters and biological materials was developed. Zinc is preconcentrated from acidic medium (pH 3.0) as its thiocyanate complex onto a polyurethane foam mini-column placed in the loop of a four-way valve. The elution step is performed with a stream of 30% acetone in 2% HNO3 and the zinc displaced is introduced directly into the nebuliser of a flame atomic absorption spectrometer. The system was operated under two different preconcentration times since the concentration of zinc in the two types of samples was different. For biological samples, 1 min was used and a detection limit of 3.0 mu g L-1 was achieved with a throughput of 40 samples per hour. At a concentration of 20 mu g L-1, an RSD of 2.5% was obtained. For the determination of zinc in natural water, a preconcentration time of 3 min was employed and a detection limit of 0.85 mu g L-1 was obtained. In this case, the system was slower with a throughput of 17 samples per hour. The RSD at 10 mu g L-1 was 6.0%. The continuous flow system was applied to the analysis of several biological reference materials and natural water samples.
引用
收藏
页码:1749 / 1753
页数:5
相关论文
共 29 条
[1]   ABSORPTION BY POLYURETHANE FOAMS - NEW METHOD OF SEPARATION [J].
BOWEN, HJM .
JOURNAL OF THE CHEMICAL SOCIETY A -INORGANIC PHYSICAL THEORETICAL, 1970, (07) :1082-&
[2]   CHROMOFOAMS QUALITATIVE AND SEMI-QUANTITATIVE TESTS WITH CHROMOGENIC ORGANIC REAGENTS IMMOBILIZED IN PLASTICIZED OPEN-CELL POLYURETHANE FOAMS [J].
BRAUN, T ;
FARAG, AB .
ANALYTICA CHIMICA ACTA, 1974, 73 (02) :301-309
[3]   SPECTROPHOTOMETRIC DETERMINATION OF TRACES OF COBALT IN WATER AFTER PRECONCENTRATION ON REAGENT-LOADED POLYURETHANE FOAMS [J].
BRAUN, T ;
ABBAS, MN .
ANALYTICA CHIMICA ACTA, 1980, 119 (01) :113-119
[4]   PLASTICIZED OPEN-CELL POLYURETHANE FOAM AS A UNIVERSAL MATRIX FOR ORGANIC REAGENTS IN TRACE-ELEMENT PRECONCENTRATION .1. COLLECTION OF SILVER TRACES ON DITHIZONE FOAM [J].
BRAUN, T ;
FARAG, AB .
ANALYTICA CHIMICA ACTA, 1974, 69 (01) :85-96
[5]   UNLOADED POLYURETHANE FOAMS AS SOLID EXTRACTANTS FOR SOME METAL THIOCYANATE COMPLEXES FROM AQUEOUS-SOLUTION [J].
BRAUN, T ;
ABBAS, MN .
ANALYTICA CHIMICA ACTA, 1982, 134 (JAN) :321-326
[6]   POLYURETHANE FOAMS AND MICROSPHERES IN ANALYTICAL-CHEMISTRY - IMPROVED LIQUID-SOLID, GAS-SOLID AND LIQUID-LIQUID CONTACT VIA A NEW GEOMETRY OF SOLID-PHASE [J].
BRAUN, T ;
FARAG, AB .
ANALYTICA CHIMICA ACTA, 1978, 99 (01) :1-36
[7]   CELLULAR AND FOAMED PLASTICS AS SEPARATION MEDIA - NEW GEOMETRICAL FORM OF SOLID-PHASE IN ANALYTICAL LIQUID-SOLID CONTACT [J].
BRAUN, T ;
FARAG, AB .
TALANTA, 1975, 22 (09) :699-705
[8]  
Braun T., 1985, POLYURETHANE FOAM SO
[9]   DETERMINATION OF ZINC AND CADMIUM BY FLOW-INJECTION ANALYSIS AND CHEMI-LUMINESCENCE [J].
BURGUERA, JL ;
BURGUERA, M ;
TOWNSHEND, A .
ANALYTICA CHIMICA ACTA, 1981, 127 (JUN) :199-201
[10]   Selective determination of cobalt using polyurethane foam and 2-(2-benzothiazolylazo)-2-p-cresol as a spectrophotometric reagent [J].
Carvalho, MS ;
Fraga, ICS ;
Neto, KCM ;
Silva, EQ .
TALANTA, 1996, 43 (10) :1675-1680