Micromachined, Polyimide-Based Devices for Flexible Neural Interfaces

被引:279
作者
Stieglitz, Thomas [1 ]
Beutel, Hansjoerg [1 ]
Schuettler, Martin [1 ]
Meyer, J. -Uwe [1 ]
机构
[1] Fraunhofer Inst Biomed Engn, Sensor Syst Microsyst Dept, D-66386 Sankt Ingbert, Germany
关键词
neural prostheses; MEMS; polyimide; electrical stimulation; implant;
D O I
10.1023/A:1009955222114
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Micromachining technologies were established to fabricate microelectrode arrays and devices for interfacing parts of the central or peripheral nervous system in case of neuronal disorders. The devices were part of a neural prosthesis that allows simultaneous multichannel recording and multisite stimulation of neurons. Overcoming the brittle mechanics of silicon, we established a process technology to fabricate light-weighted and highly flexible polyimide based devices. Concerning the challenging housing demands close to the nerve to prevent mechanical induced nerve traumatization, we integrated interconnects to decouple the nerve interface from plugs and signal processing electronics. Hybrid integration with a new assembling technique D the MicroFlex interconnection (MFI)-has been applied for the connection of the flexible microsystems to silicon microelectronics. In this paper, we present different shapes and applications of the flexible electrodes: sieve electrodes for regeneration studies, cuff electrodes for interfacing peripheral nerves, and a retina implant for ganglion cell stimulation. The discussion is focused on electrode and material properties and the hybrid assembly of a fully implantable neural prosthesis.
引用
收藏
页码:283 / 294
页数:12
相关论文
共 30 条
[1]   A MICROMACHINED SILICON SIEVE ELECTRODE FOR NERVE REGENERATION APPLICATIONS [J].
AKIN, T ;
NAJAFI, K ;
SMOKE, RH ;
BRADLEY, RM .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (04) :305-313
[2]   A modular micromachined high-density connector system for biomedical applications [J].
Akin, T ;
Ziaie, B ;
Nikles, SA ;
Najafi, K .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1999, 46 (04) :471-480
[3]   Versatile 'microflex'-based interconnection technique [J].
Beutel, H ;
Stieglitz, T ;
Meyer, JU .
SMART STRUCTURES AND MATERIALS 1998: SMART ELECTRONICS AND MEMS, 1998, 3328 :174-182
[4]  
Beutel H., 2000, P IEEE 2 INT WORKSH, P9
[5]  
BOLZ A, 1995, BEDEUTUNG PHASENGREN
[6]   Multiple site silicon-based probes for chronic recordings in freely moving rats:: implantation, recording and histological verification [J].
Bragin, A ;
Hetke, J ;
Wilson, CL ;
Anderson, DJ ;
Engel, J ;
Buzsáki, G .
JOURNAL OF NEUROSCIENCE METHODS, 2000, 98 (01) :77-82
[7]   SACRAL ANTERIOR ROOT STIMULATORS FOR BLADDER CONTROL IN PARAPLEGIA - THE 1ST 50 CASES [J].
BRINDLEY, GS ;
POLKEY, CE ;
RUSHTON, DN ;
CARDOZO, L .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1986, 49 (10) :1104-1114
[8]   Artificial vision for the blind by connecting a television camera to the visual cortex [J].
Dobelle, WH .
ASAIO JOURNAL, 2000, 46 (01) :3-9
[9]   BIOCOMPATIBILITY OF A SILICON BASED PERIPHERAL-NERVE ELECTRODE [J].
EDELL, DJ ;
CHURCHILL, JN ;
GOURLEY, IM .
BIOMATERIALS MEDICAL DEVICES AND ARTIFICIAL ORGANS, 1982, 10 (02) :103-122
[10]   A PERIPHERAL-NERVE INFORMATION TRANSDUCER FOR AMPUTEES - LONG-TERM MULTICHANNEL RECORDINGS FROM RABBIT PERIPHERAL-NERVES [J].
EDELL, DJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (02) :203-214