Compact CT/SPECT small-animal imaging system

被引:53
作者
Kastis, GA [1 ]
Furenlid, LR
Wilson, DW
Peterson, TE
Barber, HB
Barrett, HH
机构
[1] Univ Arizona, Div Nucl Med, Dept Radiol, Tucson, AZ 85724 USA
[2] Vanderbilt Univ, Inst Imaging Sci, Nashville, TN 37232 USA
[3] Univ Arizona, Ctr Opt Sci, Tucson, AZ 85721 USA
关键词
biomedical imaging; cadmium zinc telluride (CdZnTe); multimodality imaging; single-photon emission computed tomography; small-animal imaging;
D O I
10.1109/TNS.2004.823337
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We have developed a dual-modality CT/SPECT imaging system for small-animal imaging applications. The X-ray system comprises a commercially available micro-focus X-ray tube and a CCD-based X-ray camera. X-ray transmission measurements are performed based on cone-beam geometry. Individual projections are acquired by rotating the animal about a vertical axis in front of the CCD detector. A high-resolution CT image is obtained after reconstruction using an ordered subsets-expectation maximization (OS-EM) reconstruction algorithm. The SPECT system utilizes a compact semiconductor camera module previously developed in our group. The module is mounted perpendicular to the X-ray tube/CCD combination. It consists of a 64 x 64 pixellated CdZnTe detector and a parallel-hole tungsten collimator. The field of view is 1 square inch. Planar projections for SPECT reconstruction are obtained by rotating the animal in front of the detector. Gamma-ray and X-ray images are presented of phantoms and mice. Procedures for merging the anatomical and functional images are discussed.
引用
收藏
页码:63 / 67
页数:5
相关论文
共 22 条
[1]  
BALZER SJ, 2002, THESIS U ARIZONA TUC
[2]   MicroPET: A high resolution PET scanner for imaging small animals [J].
Cherry, SR ;
Shao, Y ;
Silverman, RW ;
Meadors, K ;
Siegel, S ;
Chatziioannou, A ;
Young, JW ;
Jones, WF ;
Moyers, JC ;
Newport, D ;
Boutefnouchet, A ;
Farquhar, TH ;
Andreaco, M ;
Paulus, MJ ;
Binkley, DM ;
Nutt, R ;
Phelps, ME .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1161-1166
[3]  
*EASTM KOD IM SENS, 2001, KAF 1001E PERF SPEC
[4]  
FURENLID LR, 2002, P 2002 IEEE NUCL SCI, V3, P1375
[5]  
GOERTZEN AL, 2002, MOL IMAGING BIOL, V4, P20
[6]  
Iwata K, 2002, J NUCL MED, V43, p10P
[7]  
Iwata K, 1999, IEEE NUCL SCI CONF R, P1608, DOI 10.1109/NSSMIC.1999.842874
[8]  
IWATA K, 2002, MOL IMAG BIOL S1, V4, pS21
[9]  
IWATA K, 2000, P 2000 IEEE NUCL SCI, V3, P16
[10]   A 3D HIDAC-PET camera with sub-millimetre resolution for imaging small animals [J].
Jeavons, AP ;
Chandler, RA ;
Dettmar, CAR .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (03) :468-473