摘要
Objectives: We have previously shown the benefits of dynamic fixation over conventional static fixation of bioprosthetic valves, In an attempt to increase the durability of bioprosthetic heart valves, we explored the benefit of low-concentration glutaraldehyde dynamic fixation, Methods: Pig aortic valves obtained fresh from the abattoir and excised with the entire root were dynamically fixed in glutaraldehyde phosphate buffer solutions varying in concentration from 0.05% to 2.5%., Denaturation temperatures were measured and mechanical testing was performed at low (3 mm/sec) to high physiologic rates (30 mm/sec) at 37 degrees C in isotonic modified Hanks solution, Results: When fixed dynamically in 0.05% glutaraldehyde solution for 24 hours, the tissue reached a degree of cross-linking (denaturation temperature = 82.8 degrees +/- 0.6 degrees C) significantly higher than that obtained for 0.05% static fixation (denaturation temperature = 793 degrees +/- 0.9 degrees C) (p < 0.05) but similar to that for conventional static fixation in 0.5% glutaraldehyde solution (denaturation temperature = 83.5 degrees +/- 0.3 degrees C). After fixation in low-concentration glutaraldehyde (0.05%), final relaxation slopes and moduli in the circumferential direction were significantly higher than those for the statically fixed tissue but similar to those for the fresh tissue, However, both dynamic and static fixation had the effect of increasing tissue extensibility to similar extents in both directions, irrespective of glutaraldehyde concentration. Conclusion: Dynamic glutaraldehyde fixation of a porcine aortic valve at lower concentrations resulted in a better degree of cross-linking and a material with biomechanical properties that more closely mimic those of natural heart valve tissue.
引用
收藏
页码:302 / 310
页数:9
相关论文
共 18 条
[1]
ANGELL WW, 1979, TISSUE HEART VALVES, P89
[2]
INFLUENCE OF FIXATION CONDITIONS ON THE PERFORMANCE OF GLUTARALDEHYDE-TREATED PORCINE AORTIC VALVES - TOWARDS A MORE SCIENTIFIC BASIS [J].
BROOM, ND ;
THOMSON, FJ .
THORAX, 1979, 34 (02) :166-176
[3]
5 TO 8-YEAR FOLLOW-UP OF PATIENTS UNDERGOING PORCINE HEART-VALVE REPLACEMENT [J].
COHN, LH ;
MUDGE, GH ;
PRATTER, F ;
COLLINS, JJ .
NEW ENGLAND JOURNAL OF MEDICINE, 1981, 304 (05) :258-262
[4]
COHN LH, 1979, TISSUE HEART VALVES, P173
[5]
David H, 1994, ASAIO J, V40, P206, DOI 10.1097/00002480-199404000-00015
[6]
Dynamic glutaraldehyde fixation of a porcine aortic valve xenograft .1. Effect of fixation conditions on the final tissue viscoelastic properties [J].
Duncan, AC ;
Boughner, D ;
Vesely, I .
BIOMATERIALS, 1996, 17 (19) :1849-1856
[7]
FUNG FC, 1993, BIOMECHANICS MECH PR, P242
[8]
IMAMURA E, 1984, J THORAC CARDIOV SUR, V88, P114
[9]
THE GLUTARALDEHYDE-STABILIZED PORCINE AORTIC-VALVE XENOGRAFT .1. TENSILE VISCOELASTIC PROPERTIES OF THE FRESH LEAFLET MATERIAL [J].
LEE, JM ;
COURTMAN, DW ;
BOUGHNER, DR .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1984, 18 (01) :61-77
[10]
THE GLUTARALDEHYDE-STABILIZED PORCINE AORTIC-VALVE XENOGRAFT .2. EFFECT OF FIXATION WITH OR WITHOUT PRESSURE ON THE TENSILE VISCOELASTIC PROPERTIES OF THE LEAFLET MATERIAL [J].
LEE, JM ;
BOUGHNER, DR ;
COURTMAN, DW .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1984, 18 (01) :79-98