Toward optimizing a preshaped catheter and system parameters to achieve single lead DDD pacing

被引:8
作者
Brownlee, RR [1 ]
Swindle, MM [1 ]
Bertolet, R [1 ]
Neff, P [1 ]
机构
[1] MED UNIV S CAROLINA,DEPT COMPARAT MED,CHARLESTON,SC 29425
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1997年 / 20卷 / 05期
关键词
single pass DDD; atrial sensing; Hanford miniature swine;
D O I
10.1111/j.1540-8159.1997.tb06790.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
P wave electrogram amplitudes and atrial stimulation thresholds were determined in eight Hanford miniature swine using a preshaped catheter with an ''S'' curve in the SVC, and a major lobe in the atrium to enhance electrode contact with the atrial wall. The catheter was designed for pacing and sensing in the DDD mode. P wave amplitudes were also ascertained with two commercially available VDD leads and compared to the data from the experimental catheters, The preshaped catheter used two 6-mm(2) platinum iridium atrial electrodes with a 7-mm separation. Both atrial electrodes are on the same side of the catheter, facing outward on the major atrial lobe formed in the catheter. The P wave amplitudes were tested only in the differential bipolar configuration. For the eight preshaped catheters, the mean was 6.6 +/- 3.8 mV while for the conventional leads it was 2.9 +/- 1.6 mV. The mean atrial stimulation thresholds ranged from 1.1 +/- 0.2 V to 2.3 +/- 1.2 V, with still lower thresholds of 0.9-1.3 V when using the parallel unipolar atrial electrode configuration, in which both parts of the bipolar atrial electrode are configured as a unipolar electrode. The data suggest that bipolar stimulation may be effective if sequential reverse polarity pulses are used to achieve cathodal stimulation from each electrode of the bipolar pair, on a beat-to-beat basis.
引用
收藏
页码:1354 / 1358
页数:5
相关论文
共 9 条
[1]   TOWARD OPTIMIZING THE DETECTION OF ATRIAL DEPOLARIZATION WITH FLOATING BIPOLAR ELECTRODES [J].
BROWNLEE, RR .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (03) :431-442
[2]  
BURES AR, 1975, Patent No. 3865118
[3]  
LAJOS TZ, 1977, Patent No. 4057067
[4]  
LAU CP, 1995, PACE, V18, P1820
[5]   EXPERIENCE AND IMPLANTATION TECHNIQUES WITH A NEW SINGLE-PASS LEAD VDD PACING SYSTEM [J].
LONGO, E ;
CATRINI, V .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (07) :927-936
[6]  
REENSTIERNA B, 1983, Patent No. 4401126
[7]  
Stokes K, 1985, MODERN CARDIAC PACIN, P33
[8]  
Swindle M.M., 1992, Swine as models in biomedical research
[9]  
VARIALLE P, 1990, PACE, V13, P757