Accuracy of radial artery blood pressure determination with the Vasotrac™

被引:36
作者
Belani, KG
Buckley, JJ
Poliac, MO
机构
[1] Univ Minnesota, Dept Anesthesiol, Sch Med, Minneapolis, MN 55455 USA
[2] Medwave Inc, Arden Hills, MN 55112 USA
来源
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE | 1999年 / 46卷 / 05期
关键词
D O I
10.1007/BF03012951
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Purpose: To evaluate the accuracy of a new non-invasive method (Vasotrac(TM)) to measure blood pressure (BP) with accompanying arterial wave-form and pulse-rate display when compared with BP and waveform measured invasively. Methods: Healthy volunteers (n=53) served as subjects far the study. Blood, pressures and waveforms obtained via a radial artery catheter (IABP) were compared with non-invasive measurements obtained every 12-15 beats by the new system (Vasotrac BP) from the opposite radial artery. In a sub-group of volunteers (n=11), BP was acutely increased and decreased with isoproterenol, phenylephrine and sodium nitroprusside; Data were analyzed by determining correlation and agreement between the two systems of measurement. Waveforms obtained by the two systems were qualitatively examined. Results: Non-invasive BP measured every 12-15 beats by the Vasotrac correlated with IABP (systolic r(2) = 0.89; diastolic r(2) = 0.88; mean r(2) = 0.94). The actual values obtained by the two methods agreed closely with > 90% of readings being within 2SDs when plotted by the Bland Altman method. This was also true during vasoactive drug infusion when BP changed acutely and swiftly. During this dynamic period, Vasotrac BP accurately tracked the changes in IABP with correlations (systolic r(2)= 0.82; diastolic r(2) = 0.89; mean r(2) = 0.95) and close agreement (> 90% of readings were within 2 SDs in the Bland Altman plot). Waveforms displayed by the two systems were qualitatively very similar. Pulse rates measured by the two systems were identical. Conclusions: The Vasotrac system displayed an arterial waveform which was similar to that obtained directly and measured BP and pulse rate accurately It should be a convenient device to measure BP continually in a non-invasive fashion.
引用
收藏
页码:488 / 496
页数:9
相关论文
共 17 条
  • [1] *AM SOC AN, 1997, ASA STAND GUID STAT, P2
  • [2] *ASS ADV MED INSTR, 1986, AM NAT STAND INT PER, P31
  • [3] Bause G S, 1986, J Clin Monit, V2, P262, DOI 10.1007/BF02851174
  • [4] BETTS EK, 1981, ANESTHESIOLOGY, V55, P717
  • [5] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [6] EICHHORN JH, 1986, JAMA-J AM MED ASSOC, V256, P1017
  • [7] GORBACK M S, 1990, Biomedical Instrumentation and Technology, V24, P371
  • [8] Gravenstein J S, 1985, J Clin Monit, V1, P120, DOI 10.1007/BF02832199
  • [9] AN ASSESSMENT OF THE DINAMAP 845
    HUTTON, P
    DYE, J
    PRYSROBERTS, C
    [J]. ANAESTHESIA, 1984, 39 (03) : 261 - 267
  • [10] ARTERIAL TONOMETRY FOR NONINVASIVE, CONTINUOUS BLOOD-PRESSURE MONITORING DURING ANESTHESIA
    KEMMOTSU, O
    UEDA, M
    OTSUKA, H
    YAMAMURA, T
    WINTER, DC
    ECKERLE, JS
    [J]. ANESTHESIOLOGY, 1991, 75 (02) : 333 - 340