High performance amorphous silicon image sensor for x-ray diagnostic medical imaging applications

被引:26
作者
Weisfield, RL [1 ]
Hartney, M [1 ]
Schneider, R [1 ]
Aflatooni, K [1 ]
Lujan, R [1 ]
机构
[1] Xerox New Enterprise Co, DpiX, Palo Alto, CA 94304 USA
来源
MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2 | 1999年 / 3659卷
关键词
a-Si; TFT; photodiode; imager; sensor; x-ray; array;
D O I
10.1117/12.349505
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Following our previous report(1) concerning the development of a 127 mu m resolution, 7.4 million pixel, 30 x 40 cm(2) active area, flat panel amorphous Silicon (a-Si) x-ray image sensor, this paper describes enhancements in image sensor performance in the areas of image lag, linearity, sensitivity, and electronic noise. New process improvements in fabricating a-Si thin film transistor (TFT)/photodiode arrays have reduced first-frame image lag to less than 2%, and uniformity in photoresponse to < 5% over the entire 30 x 40 cm2 active area. Detailed analysis of image lag vs, time and x-ray dose will be discussed. An improved charge amplifier has been introduced to suppress image cross-talk artifacts caused by charge amplifier saturation, and system linearity has been optimized to eliminate banding effects among charge amplifiers. Preliminary sensitivity improvements through the deposition of CsI(TI) directly on the arrays are reported, as well as overall imaging characteristics of this improved image sensor.
引用
收藏
页码:307 / 317
页数:11
相关论文
共 9 条
[1]  
ANTONUK LE, 1998, SPIE, V3336, P2
[2]   Technical and clinical results of an experimental Flat Dynamic (digital) X-ray image Detector (FDXD) system with real-time corrections [J].
Bruijns, TJC ;
Alving, PL ;
Baker, EL ;
Bury, R ;
Cowen, AR ;
Jung, N ;
Luijendijk, HA ;
Meulenbrugge, HJ ;
Stouten, HJ .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :33-44
[3]  
COLBETH RE, 1998, SPIE, V3336, P376
[4]  
LEE DL, 1995, P SOC PHOTO-OPT INS, V2432, P237, DOI 10.1117/12.208342
[5]   Flat panel detector for digital radiology using active matrix readout of amorphous selenium [J].
Rowlands, JA ;
Zhao, W ;
Blevis, I ;
Pang, GD ;
Ji, WG ;
Germann, S ;
Kasap, SO ;
Waechter, D ;
Huang, ZS .
PHYSICS OF MEDICAL IMAGING - MEDICAL IMAGING 1997, 1997, 3032 :97-108
[6]  
STREET R, 1998, SPIE, V3336, P24
[7]   A PHYSICAL INTERPRETATION OF DISPERSIVE TRANSPORT IN DISORDERED SEMICONDUCTORS [J].
TIEDJE, T ;
ROSE, A .
SOLID STATE COMMUNICATIONS, 1981, 37 (01) :49-52
[8]   Development of a selenium-based flat-panel detector for real-time radiography and fluoroscopy [J].
Tsukamoto, A ;
Yamada, S ;
Tomisaki, T ;
Tanaka, M ;
Sakaguchi, T ;
Asahina, H ;
Nishiki, M .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :388-395
[9]  
WEISFIELD RL, 1998, SPIE, V3336, P444