Effect of two-level pedicle-screw fixation with different rod materials on lumbar spine: A finite element study

被引:75
作者
Biswas, Jayanta Kumar [1 ]
Rana, Masud [2 ]
Majumder, Santanu [1 ]
Karmakar, Santanu Kumar [2 ]
Roychowdhury, Amit [1 ]
机构
[1] Indian Inst Engn Sci & Technol, Dept Aerosp Engn & Appl Mech, Sibpur 711103, Howrah, India
[2] Indian Inst Engn Sci & Technol, Dept Mech Engn, Sibpur 711103, Howrah, India
关键词
IMPLANT; BONE; DISC; POLYETHERETHERKETONE; DIMENSIONS; ROTATION; TITANIUM; SEGMENT; FUSION; TISSUE;
D O I
10.1016/j.jos.2017.10.009
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
Background: Pedicle-screw-rod fixation system is very popular surgical remedy for degenerative disc disease. It is important to observe load vs. spinal motion characteristic for better understanding of clinical problems and treatment of spinal instability associated with low-back pain. Objective: The objective of this study is to understand the effect [range of motion (ROM) and intervertebral foramen height] of pedicle-screw fixation with three rod materials on lumbar spine under three physiological loading conditions. Method: A three-dimensional finite element (FE) model of lumbar to sacrum (L1-S) vertebrae with pedicle-screw-rod fixation at L3-L5 level is developed. Three rod materials [titanium alloy (Ti-6Al-4V), ultra-high molecular weight poly ethylene (UHMWPE) and poly-ether-ether-ketone (PEEK)] are used for two-level fixation and the FE models are simulated for axial rotation, lateral bending and flexion-extension under +/- 10 Nm moment and 500 N compressive load and compared with the intact (natural) model. Result & discussion: For axial rotation, lateral bending and flexion, ROM increased 2.8, 4.5 and 1.83 times respectively for UHMWPE, and 3.7, 7.2 and 2.15 times respectively for PEEK in comparison to Ti-6Al-4V. As ROM is 49, 29 and 31% of the intact model during axial rotation, lateral bending and flexion respectively, PEEK rod produced better motion flexibility than Ti-6Al-4V and UHMWPE rod. Foramen height increased insignificantly by 2.21% for the PEEK rod with respect to the intact spine during flexion. For the PEEK rod, maximum stress of 40 MPa during axial rotation is much below the yield stress of 98 MPa. Conclusion: Ti-6Al-4V pedicle-screw-rod fixation system highly restricted the ROM of the spine, which is improved by using UHMWPE and PEEK, having lower stiffness. The foramen height did not vary significantly for any implant materials. In terms of ROM and maximum stress, PEEK rod may be considered for a better implant design to get better ROM and thus mobility. (c) 2017 The Japanese Orthopaedic Association. Published by Elsevier B.V. All rights reserved.
引用
收藏
页码:258 / 265
页数:8
相关论文
共 36 条
[1]
Load sharing in lumbar spinal segment as a function of location of center of rotation [J].
Alapan, Yunus ;
Sezer, Semih ;
Demir, Cihan ;
Kaner, Tuncay ;
Inceoglu, Serkan .
JOURNAL OF NEUROSURGERY-SPINE, 2014, 20 (05) :542-549
[2]
Bilateral pedicle screw fixation provides superior biomechanical stability in transforaminal lumbar interbody fusion: a finite element study [J].
Ambati, Divya V. ;
Wright, Edward K., Jr. ;
Lehman, Ronald A., Jr. ;
Kang, Daniel G. ;
Wagner, Scott C. ;
Dmitriev, Anton E. .
SPINE JOURNAL, 2015, 15 (08) :1812-1822
[3]
Comparison of trunk muscle forces and spinal loads estimated by two biomechanical models [J].
Arjmand, N. ;
Gagnon, D. ;
Plamondon, A. ;
Shirazi-Adl, A. ;
Lariviere, C. .
CLINICAL BIOMECHANICS, 2009, 24 (07) :533-541
[4]
Biswas Jayanta, 2014, J Long Term Eff Med Implants, V24, P99
[5]
Chang TK., 2013, FORMOSAN J MUSCULOSK, V4, P42
[6]
Biomechanical analysis of different types of pedicle screw augmentation: A cadaveric and synthetic bone sample study of instrumented vertebral specimens [J].
Chao, Kuo-Hua ;
Lai, Yu-Shu ;
Chen, Wen-Chuan ;
Chang, Chia-Ming ;
McClean, Colin J. ;
Fan, Chang-Yuan ;
Chang, Chia-Hao ;
Lin, Leou-Chyr ;
Cheng, Cheng-Kung .
MEDICAL ENGINEERING & PHYSICS, 2013, 35 (10) :1506-1512
[7]
Efficacy of anterior cervical fusion: Comparison of titanium cages, polyetheretherketone (PEEK) cages and autogenous bone grafts [J].
Chou, Yu-Cheng ;
Chen, Der-Cherng ;
Hsieh, Wanhua Annie ;
Chen, Wu-Fu ;
Yen, Pao-Sheng ;
Harnod, Tomor ;
Chiou, Tsung-Lang ;
Chang, Yuh-Lin ;
Su, Chain-Fa ;
Lin, Shinn-Zong ;
Chen, Shin-Yuan .
JOURNAL OF CLINICAL NEUROSCIENCE, 2008, 15 (11) :1240-1245
[8]
A comparison of stress distributions for different surgical procedures, screw dimensions and orientations for a Temporomandibular joint implant [J].
Chowdhury, Amit Roy ;
Kashi, Ajay ;
Saha, Subrata .
JOURNAL OF BIOMECHANICS, 2011, 44 (14) :2584-2587
[9]
Cosar M, 2008, EUR J GEN MED, V5, P147
[10]
Direct measurement of intervertebral disc maximum shear strain in six degrees of freedom: Motions that place disc tissue at risk of injury [J].
Costi, J. J. ;
Stokes, I. A. ;
Gardner-Morse, M. ;
Laible, J. P. ;
Scoffone, H. M. ;
Iatridis, J. C. .
JOURNAL OF BIOMECHANICS, 2007, 40 (11) :2457-2466