Modeling and control of a brushless DC axial flow ventricular assist device

被引:66
作者
Giridharan, CA
Skliar, M [1 ]
Olsen, DB
Pantalos, GM
机构
[1] Univ Utah, Dept Chem & Fuels Engn, Salt Lake City, UT 84112 USA
[2] Utah Artificial Heart Inst, Salt Lake City, UT USA
[3] Univ Louisville, Jewish Hosp, Heart & Lung Inst, Louisville, KY 40292 USA
关键词
D O I
10.1097/00002480-200205000-00013
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This article presents an integrated model of the human circulatory system that incorporates circulatory support by a brushless DC axial flow ventricular assist device (VAD), and a feedback VAD controller designed to maintain physiologically sufficient perfusion. The developed integrated model combines a network type model of the circulatory system with a nonlinear dynamic model of the brushless DC pump. We show that maintaining a reference differential pressure between the left ventricle and aorta leads to adequate perfusion for different pathologic cases, ranging from normal heart to left heart asystole, and widely varying physical activity scenarios from rest to exercise.
引用
收藏
页码:272 / 289
页数:18
相关论文
共 16 条
[1]  
ARTHUR CG, 1981, TXB MED PHYSL
[2]  
BENHAM RD, 1966, SIMULATION, V7, P324
[3]  
CHOI S, 1997, AM CONTR C ALB NM P
[4]   HABITUATING CONTROL STRATEGIES FOR PROCESS-CONTROL [J].
HENSON, MA ;
OGUNNAIKE, BA ;
SCHWABER, JS .
AICHE JOURNAL, 1995, 41 (03) :604-618
[5]   THE BARORECEPTOR REFLEX - A BIOLOGICAL-CONTROL SYSTEM WITH APPLICATIONS IN CHEMICAL PROCESS-CONTROL [J].
HENSON, MA ;
OGUNNAIKE, BA ;
SCHWABER, JS ;
DOYLE, FJ .
INDUSTRIAL & ENGINEERING CHEMISTRY RESEARCH, 1994, 33 (10) :2453-2466
[6]   SYSTEM CONSIDERATIONS FAVORING ROTARY ARTIFICIAL HEARTS WITH BLOOD-IMMERSED BEARINGS [J].
JARVIK, RK .
ARTIFICIAL ORGANS, 1995, 19 (07) :565-570
[7]  
KNUDSON KPC, 1998, IN VITRO CHARACTERIS
[8]  
Konishi H, 1994, ASAIO J, V40, pM299, DOI 10.1097/00002480-199407000-00012
[9]  
MCLEOD J, 1972, SIMULATION, P152
[10]   Development and initial testing of a permanently implantable centrifugal pump [J].
Nakazawa, T ;
Takami, Y ;
Benkowski, R ;
Ohtsubo, S ;
Yukio, O ;
Tayama, E ;
Ohtsuka, G ;
Niimi, Y ;
Glueck, J ;
Sueoka, A ;
Schmallegger, H ;
Schima, H ;
Wolner, E ;
Nose, Y .
ARTIFICIAL ORGANS, 1997, 21 (07) :597-601