Determination of total organic carbon in water by thermal combustion-ion chromatography

被引:19
作者
Fung, YS
Wu, ZC
Dao, KL
机构
[1] Department of Chemistry, University of Hong Kong, Hong Kong, Pokfulam Road
关键词
D O I
10.1021/ac951146x
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
sensitive method for determining total organic carbon at microgram per liter levels in industrial, environmental, and drinking waters by thermal combustion ion chromatography was developed using tube furnace and readily accessable HPLC equipment. To achieve complete oxidation, persulfate (0.25%) was added to oxidize nonvolatile organic compounds in solution and cupric oxide heated at 900 degrees C to convert volatile organic compounds to CO2, which was scrubbed in a 20 mt solution of 50 mM KOH with 10 drops of butanol added. The carbonate anion obtained was determined by nonsuppressed ion chromatography using 0.6 mM potassium hydrogen phthalate (KHP) as the eluent. Both surfactants and volatile and nonvolatile organic compounds commonly found in environmental waters give highly repeatable recoveries close to 100%. The detection limit (S/N = 2) and linear range for a 1 L water sample are 2 mu g of C L(-1) and 10-2500 mu g of C L-l, respectively, and they can be adjusted using samples ranging from 100 mt to 2 L. Good repeatablity (RSD less than 10%) and close to 100% recoveries were obtained for IMP added to real samples such as deionized, mineral, tap, and river water and seawater, Compared with the ASTM D2579 method, the method developed is 3 orders of magnitude more sensitive, more accurate, and reliable in determining samples with low total organic carbon values and more flexible in adjusting the linear range and sensitivity using variable sample sizes.
引用
收藏
页码:2186 / 2190
页数:5
相关论文
共 27 条
[1]  
*AM PUBL HLTH ASS, 1989, 5310C AM PUBL HLTH A, P5
[2]  
*ASTM, 1985, TEST METH TOT ORG CA
[3]   A TEST OF THE ACCURACY OF FRESH-WATER DOC MEASUREMENTS BY HIGH-TEMPERATURE CATALYTIC-OXIDATION AND UV-PROMOTED PERSULFATE OXIDATION [J].
BENNER, R ;
HEDGES, JI .
MARINE CHEMISTRY, 1993, 41 (1-3) :161-165
[4]   ION-CHROMATOGRAPHIC BEHAVIOR OF CARBONATE SPECIES [J].
BRANDT, G ;
KETTRUP, A .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1985, 320 (05) :485-489
[5]   ANALYSIS OF ANIONS BY ION CHROMATOGRAPHY USING ULTRAVIOLET DETECTION [J].
COCHRANE, RA ;
HILLMAN, DE .
JOURNAL OF CHROMATOGRAPHY, 1982, 241 (02) :392-394
[6]  
EDWARDS RT, 1992, ANAL CHIM ACTA, V26, P287
[7]   FLOW-INJECTION ANALYSIS OF ORGANIC AND INORGANIC CARBON IN THE LOW-PPB RANGE [J].
HUBER, SA ;
FRIMMEL, FH .
ANALYTICAL CHEMISTRY, 1991, 63 (19) :2122-2130
[8]   THE OCCURRENCE AND ORIGIN OF SYSTEM PEAKS IN NON-SUPPRESSED ION CHROMATOGRAPHY OF INORGANIC ANIONS WITH INDIRECT ULTRAVIOLET-ABSORPTION DETECTION [J].
JACKSON, PE ;
HADDAD, PR .
JOURNAL OF CHROMATOGRAPHY, 1985, 346 (OCT) :125-137
[9]   CONTINUOUS RECORDING OF CONCENTRATION OF ORGANIC MATTER IN WASTE WATER [J].
KIESELBACH, R .
ANALYTICAL CHEMISTRY, 1954, 26 (08) :1312-1318
[10]   ION CHROMATOGRAPHIC PROCEDURE FOR BICARBONATE DETERMINATION IN BIOLOGICAL-FLUIDS [J].
KRELING, JR ;
DEZWAAN, J .
ANALYTICAL CHEMISTRY, 1986, 58 (14) :3028-3031